Input Functions
In addition to the source functions, the 8600 combines a high resolution frequency counter
that is internally disciplined to GPS reducing the requirement for separate pieces of equipment and
complicated wiring and interconnections.
Gate Settings
The measurement gate allows the user to select the duration of the measurement. Higher gate
times provide more resolution, however measurements will be slower to react to changes in frequency.
The gate time is selectable from 0.1, 1 and 10 seconds by pressing the DIGITS key on the front
panel.
Trigger Level
The trigger level sets a minimum input level at which the counter starts a measurement. For
measurements where there is the potential for distortion setting the trigger level can ensure that only
the peaks are captured and any incorrect counts caused by distortion on the waveform are ignored.
The trigger level is set by pressing the TRIGGER LEVEL key and then adjusting via the digital
control until a stable trigger is achieved.
Sensitivity
The sensitivity mode sets the hysteresis level around the trigger level. When in high sensitivity
mode the hysteresis around the trigger point is widened to enable the counter to trigger off of a wider
range of levels. When in low sensitivity mode the hysteresis around the trigger point is narrowed to
reduce the possibility of false triggering when distorted signals are being measured.
The sensitivity is set between HIGH and LOW by pressing the SENSE HIGH-LOW key. The
current sensitivity is displayed in the status display.
High Frequency Filter
The High Frequency filter assists with noise when measuring frequencies below 400 Hz by
placing a passive low pass filter in the internal signal chain to prevent high frequencies triggering the
counter.
To enable or disable the High Frequency filter press the HF FILTER key. The current filter
state is displayed in the status display as Flt Off or Flt on depending on the present status of the filter.