Page 29
Electromagnetic Immunity
The Transcend
Micro is intended
for use in the
electromagnetic
environment
specified below.
The customer or
the user of the
system should
ensure that it is
used in such an
environment.
Basic EMC standard
Immunity test levels
Electrostatic
discharge (ESD)
IEC 61000-4-2
±2, 4, 8, 15kV air
±8 kV contact
Radiated RF EM
fields
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV
100 kHz repetition frequency
Surges
Line-to-ground
IEC 61000-4-5
±1 0, 5 kV ± 1kV
Surges
Line-to-ground
IEC 61000-4-5
± 0, 5 kV, ± 1kV, ± 2kV
Voltage dips,
IEC 61000-4-11
0,5 cycle
At 0, 45, 90, 135, 180, 225, 270
and 315 degrees
0% Ut; 1 cycle and 70% Ut;
25/30 cycles
Single phase: at 0
Voltage interruptions IEC 61000-4-11
0% Ut; 250/300 cycle