Page 18
1.
Depress the tab on the right side of the filter cover to remove it from the device
2.
Remove the filter and discard. Replace with a new filter.
3.
Reattach the filter cover by inserting the left side into the device and then clicking
the right side in.
Cleaning for Multiple Users
If using the device on multiple users, perform the following steps to clean the device
before each new user:
1.
Unplug the power supply prior to cleaning.
2.
Remove and discard the bacterial/viral filter.
3.
Remove and discard the used air supply tube.
4.
Follow instructions for cleaning the exterior of the device as noted in Cleaning the
Exterior (above).
5.
Apply a new bacterial/viral filter between the hose and the mask before providing
the device to a new user.
Warning
The air supply tubing, mask assembly and bacterial/viral filters should be
discarded after each patient use using standard institutional biohazard
procedures. No attempt should be made to clean, disinfect, or sterilize these
components for multiple users. These components are for single patient use.
Manufacturer’s instructions must be followed with regard to cleaning,
disinfecting, and re-use of patient interfaces (masks).
Warning
Only use bacterial/viral filters that are commercially available and designed
for use with CPAP machines. Only use filters that are ISO 23328-1:2003 and
ISO 23328-2:2002 approved.
Follow the manufacturer’s instructions for use
of the filter. Only use those commercially available filters that do not require
recalibration of the PAP device.
Warning
Do not submerge the Transcend Micro or power supply in liquid. Do not allow
liquid or cleaning solution to enter the device.
Environmental Information
This device should be disposed of separately, not as unsorted municipal waste. To dispose
of your device, you should use appropriate collection and recycling systems available in
your region. If information on these disposal systems is needed, please contact your local
waste administration. If you require information on the collection or disposal of your
device please contact Transcend, your homecare provider or go to
MyTranscend.com/environment.
California Perchlorate Information:
The coin-cell battery within this device may contain Perchlorate Material
–
special handling
may apply. See: