Page 4-1
CHAPTER 4: SETUP MENU DESCRIPTIONS AND PROCEDURES
4.1
SETUP MENU DESCRIPTIONS
This section provides more detailed descriptions of the selections found in the Setup Menu Chart.
Factory-set defaults are shown in bold with a checkmark (
√
).
NAME/CODE DESCRIPTION CODE/VALUE
F1
Graduations
Specifies number of full-scale graduations. Value should be consis-
tent with legal requirements and environmental limits on the useful
system resolution.
10,000
√
F3
Zero Track
Band
Selects the range within which the scale will automatically zero. Note
that the scale must be in standstill to automatically zero. Selections
are in Display Divisions.
0d
0.1d
0.3d
0.5d
√
1d
3d
5d
F8
Calib. Unit
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1
2
√
depends on
model
F9
Display
Divisions
Determines the desired weight increments. Value should be consis-
tent with legal requirements.
1
2
5
√
depends on
model
F10
Decimal Pt.
Determines location of the decimal point.
0.0
0.000 0.0000
√
depends on
model
F16
Zero Calibra-
tion
Places indicator into the zero calibration routine. Scrolling down with
the
6
key one level begins the procedure.
Press
6
key to
begin sequence
F17
Span Calibra-
tion
Places indicator into the span calibration routine. Scrolling down with
the
6
key one level begins the procedure.
Press
6
key to
begin sequence
F22
Creep Track-
ing Band
Determines the creep tracking band in scale divisions.
OFF
√
10 15 20
25 30 50
F23
Creep Track-
ing Time
Determines the creep tracking time in seconds.
1 2 3 4
5
√
10 20 30