Page 6-1 

 

CHAPTER 6: CALIBRATION  

 

6.1 CALIBRATION 

OVERVIEW 

 
The scale is calibrated by following the procedures embedded in F16 (Zero) and F17 (Span) of the 
Setup Menu. Each procedure enters a value into the scale's non-volatile memory - F16 the zero 
value (deadweight) and F17 the span value (test weight). The minimum test weight that can be 
used is 1% of full-scale capacity.  
 

NOTE:

 This chapter assumes that the scale is in Setup (“F”) Menu mode. If the scale is not in 

Setup Menu mode, refer to Chapter 3 for instructions. 
 

6.2 

ZERO CALIBRATION (F16) 

 

1.  While in the Setup mode, scroll to "

F 16

", then scroll down once using the 6 key to enter zero 

calibration menu. The display will momentarily show "

C 0

" followed by a value. This value is the 

internal A/D count and can prove useful when trying to troubleshoot setup problems. 

2.  After making sure that there are no test weights on the platform, press the ZERO key to zero 

out the displayed value. 

3.  Press the SAMPLE/SAVE key to save the zero point value. The display will show "

EndC0

momentarily, then revert back up to F16. At this time, proceed to the F17 span calibration to 
complete scale calibration. 

 

6.3 

SPAN CALIBRATION (F17) 

 

1.  While in the Setup mode, scroll to "

F 17

", then scroll down once using the 6 key to enter span 

calibration menu. 

2.  The display will momentarily show "

C 1

" for the span calibration, followed by a value with one 

flashing digit. This value will be zero with the Decimal Point parameter selected in F10. Place the 
test weight on the platter. 

3.  Use the numeric keys to enter the test weight value. Please note that the decimal point is fixed so 

that if you are entering, for example, 2 lb, you need to type “2” followed by zeroes until the “2” is 
to the left of the decimal point. 

4.  After setting the exact value, press the SAMPLE/SAVE key to save the value. 

5.  If the calibration was successful, the display will show "

EndC1

" momentarily, then revert back up 

to F17.  

6.  If the calibration was 

not 

successful, one of the error messages below will appear. Take the 

indicated action to correct the problem, then perform a new calibration. 

"

Err0

" - The calibration test weight or the adjusted keyed-in weight is larger than the full 

capacity of the scale. Change the calibration test weight or check the input data. 

"

Err1

" - The calibration test weight or the adjusted keyed-in weight is smaller than 1% of the full 

capacity of the scale. Change the calibration test weight or check the input data. 

"

Err2

" - The internal resolution of the scale is not high enough to accept the calibration value.  

 

 

Summary of Contents for TC-100 Series

Page 1: ...1 0 January 10 2003 2003 Transcell Technology Inc Contents subject to change without notice Transcell Technology Inc 975 Deerfield Parkway Buffalo Grove IL 60089 Tel 847 419 9180 Fax 847 419 1515 E m...

Page 2: ...1 3 2 3 Notes on the Setup Menu 3 2 3 2 4 Exiting the Setup Menu 3 2 3 3 User A Menu 3 2 Chapter 4 Setup Menu Descriptions and Procedures 4 1 4 1 Setup Menu Descriptions 4 1 4 2 Resetting Factory Def...

Page 3: ...menu which configures the Scale to your weigh platform and the User A menu which configures the serial communication port and enables some user op tions Chapter 3 gives an overview and explains how t...

Page 4: ...ed for connection to either a PC or a serial printer It can also be connected to a simplex device such as a remote display COM1 is wired to a physical connector a D SUB9 female Refer to the TC 100 Ope...

Page 5: ...e scale shows F 1 you are in Setup Menu mode and you may release the 0 key 3 2 2 NAVIGATING IN THE SETUP MENU Use the directional keys shown in Figure 3 1 to move around in the Setup Menu Chart shown...

Page 6: ...ETUP MENU 1 Detailed descriptions of the setup menu parameters can be found in Chapter 5 of this manual 3 2 4 EXITING THE SETUP MENU 1 Power off the scale by toggling the ON OFF switch 2 Power on the...

Page 7: ...nit to be used in the calibration process Also the default unit for normal operation 1 primary unit is lb 2 primary unit is in kg 1 2 depends on model F9 Display Divisions Determines the desired weigh...

Page 8: ...while switching the scale back on You may release that key when the scale displays the corresponding message found in Table 4 1 MODEL CAPACITY GRADUATION KEY MESSAGE TC 100 2 2 lb x 0 0002 lb 1 INIT 2...

Page 9: ...Page 5 1 CHAPTER 5 USER MENU DESCRIPTIONS AND PROCEDURES 5 1 USER MENU DESCRIPTIONS Please refer to Chapter 5 of the TC 100 Operation Manual...

Page 10: ...1 While in the Setup mode scroll to F 17 then scroll down once using the 6 key to enter span calibration menu 2 The display will momentarily show C 1 for the span calibration followed by a value with...

Reviews: