Page 6-2
percent markup for the displayed rate category is OK. Repeat Step 2.
Note:
If the desired markup percentage is less than 10, you must press the ENTER
key. Otherwise, if the desired markup percentage is 10 or greater, the scale
automatically enters the value for you.
4.
Repeat Steps 2 and 3 for all rate categories listed.
6.2.3 BUZZER ENABLE/DISABLE
This function is used to enable or disable the audible tone heard when a key is pressed.
1.
Enter the Setup Menu Mode as described above and press the ENTER key until the
scale prompts ” Buzzer OK?” and displays the current status (ON/OFF) of the buzzer.
2.
Press the ENTER (Yes) key to accept the displayed buzzer status or press the CLEAR
(No) key to change the buzzer status.
Note:
If
“Yes”
is selected, the scale proceeds to Normal Operating Mode. If
“No”
is
selected, the status toggles between ON and OFF.
6.3
CALIBRATION MENU
Your SPS scale ships from the factory fully calibrated. There is no need to calibrate the scale
unless you feel that it has become inaccurate. It is a good idea to check the calibration of your
scale from time to time with a precision test weight. Transcell recommends that you perform a new
calibration on your digital scale at least once a year.
The SPS-70 scale may be calibrated using precision test weights of 4 lb and 50 lb. The SPS-150
scale may be calibrated using precision test weights of 4 lb and 100 lb. Please note that if the unit
is sealed for use in commercial applications, this calibration procedure is inhibited. See Chapter 7
for more information.
1. Enter the Calibration Mode by pressing and holding the Mexico (6) key while turning on the
scale. When the message “Calibration? Y/N” appears on the screen, you are in Calibration
Mode and you may release the Mexico (6) key.
2. Press the ENTER (Yes) key enter the calibration mode. The scale displays the message “C 0
PR ZERO”.
3. Remove all items from the scale’s platter and then press the ZERO key to zero the scale. The
scale displays the message “C 0 PR ENTER”.
4. Press the ENTER key to save the zero point value. The scale displays the message “C 4 “.
5. Place the 4 lb test weight(s) onto the scale’s platter. When the weight stabilizes, the scale
displays the message “C 4 PR ENTER”.
6. Press the ENTER (Yes) key to save the 4 lb calibration value. The SPS-70 scale displays the
message “C 50 “. The SPS-150 scale displays the message “C 100 “.
7. Place either the appropriate test weight(s) onto the scale’s platter. When the weight stabilizes,
the scale displays the message “ PR ENTER”.
8. Press the ENTER (Yes) key to save the 50 lb or 100 lb calibration value. The scale then
displays the message “Calibration End”.
9. Turn OFF the scale then turn back ON to resume normal operation.
6.4
ANALOG TO DIGITAL CONVERSION (ADC) DIAGNOSTIC MENU
This menu is reserved for Transcell technical personnel only.