TROUBLE - SHOOTING
DISPLAY CAUSE
SOLUTION
Er.1
Scoreboard could not display the actual
value.
e.g. : -
1000
or
110000
Make a tare
Check parameters of Set Up 4
Er.20
Load cell out signal is lower than 400 Hz
1) Value between 1 and 399 Hz
2) Signal = 0 Hz
- Wrong wiring.
- Power supply fault.
- Display fault
- Load cell fault
NOTE : Wrong polarity wiring may
cause destruction of some electronic
components.
To check frequency signal press
Ø
1) Increase the frequency up to 500 hz
with the "zero" potentiometer.
( page 20)
2)
- Double check wiring.
- Double check power supply.
- Contact Tractel ASS
- Contact Tractel ASS
Er.21
Frequency out put signal could not be
read by the display.
- Signal greater than 9999 Hz
- Load cell is over stressed
- Display fault
- Load cell fault
NOTE : Wrong polarity wiring may
cause destruction of some electronic
components.
- adjust "GAIN" potentiometer
- compare stress and capacity
- Contact Tractel ASS
- Contact Tractel ASS