Metalyser
®
Soils HM4000
Instruction Manual v3.0
12.
Choose the “Digest” method from the list of methods in the
Metalyser
®
. Select condition and allow the stirrer to run for 10
minutes.
13.
After 10 mins, remove the beaker from the stirrer head and let it
settle for 5 minutes.
14.
Connect a syringe filter to a 5mL syringe. Attach a short length of
tubing to the filter and use the syringe for taking 3.5ml of the
digested sample.
15.
Add the 3.5 mL sample to the sample analysis beaker and fill to
the mark just below the overflow with de-ionised water.
16.
The soil extract sample is now ready to be analysed using the
Metalyser
®
.