Metalyser
®
Soils HM4000
Instruction Manual v3.0
Example:
Overview of As (III) procedure to first plate then sensitise the electrode:
Main Menu
Choose “Test Methods”
These plating steps
deposit the plate on
the tip of the
Working Electrode
Test Methods Menu
Choose “M2&3 Conditioning”
Half-fill SAB with AU500 Au Plating Solution
After plating, check tip for yellow / gold plate
Return AU500 Au Plating Solution to bottle
These
conditioning steps
(within the metal
menus) prepare
the Working
Electrode for
analysis.
Select analysis method...
Place buffer in SAB and fill SAB with
sample
From the
Test Methods Menu
, choose
As(III),
and then
Condition Electrode
Refer to Application notes for Operating Procedures particular to each parameter.