![TR Comfortline Service Manual Download Page 53](http://html1.mh-extra.com/html/tr/comfortline/comfortline_service-manual_1146640053.webp)
Hygiene Equipment
Designed for People
53
Routine maintenance TR Comfortline
Function control of Lifting columns, Thermostatic mixer and drainage
Initialize/calibrate the lifting
columns according to instructions
in the Operating manual.
Yes No
Comment:
Action:
Is it possible to raise and lower
the bathtub with the hand
control?
Yes No
Comment:
Action:
Exercise the Thermostatic mixer,
supply the coldest mixer
temperature for 30 seconds and
the hottest mixer temperature
(Max 43°) for 30 seconds. Is the
function OK?
Yes No
Comment:
Action:
Does the safety stop for 38°C
and Maximum 43°C work OK?
Yes No
Comment:
Action:
Perform a scalding test, Shut OFF
the Cold-water supply to the tub.
The mixer shall shut OFF when
only Hot water is supplied.
Scalding function working?
Yes No
Comment:
Action:
If the mixer doesn´t shut OFF when only HOT water is supplied, the mixer or cartridge MUST BE REPLACED!