Automatic Data logging, continued…
If the
WP-84
is sending records directly to the RS232 port, the display
will look like this…
2.76mS
25.0
o
c
Sending 12:00:00
The
WP-84
will beep each time a record is sent to the RS232 port.
8. Press
to stop automatic logging.
Notes
1. The clock must be set before the
WP-84
will allow automatic logging
to start. The message “
Clock Not Set
” is displayed if the clock is
not set.
2. The Battery Saver function (section 11) is disabled while the meter is
in Automatic Data logging mode, to stop the meter switching off while
logging data. Even when the memory is full and the meter stops
logging, the Battery Saver function is still disabled. This allows the
data to be downloaded and the memory to be reset remotely.
Page 24