7.
Good Laboratory Practices (GLP)
The
WP-84
keeps a record of the date and time of the last Conductivity
and Temperature calibrations as part of GLP guidelines.
7.1
To recall GLP information on the display
1.
Select the GLP menu (
�
F4:Setup
�
F3:GLP
).
2.
Select
F1:Recall
from the menu.
3.
The instrument model, firmware version number, and instrument serial
number are displayed, along with a prompt describing how to scroll
through the GLP information. For example…
WP84
V4.0 S1234
�
:More
�
:Back
4.
Press the
key to sequentially scroll through the GLP information for
all parameters. Press the
key to scroll back to previous data. The
sequence of information displayed is shown over the page. Press
to abort at any time.
GLP display sequence…
WP84
V4.0 S1234
�
:More
�
:Back
Cond Zero 0.00uS
@ 31/12/10 12:00
k=1.00 @ 2.76mS
@ 31/12/10 12:10
Temp Offset 1.0
o
c
@ 31/12/10 12:30
Exit
7.2
Failed Calibration
If calibration has failed, the GLP function will reset the date and time to
zero. The
WP-84
still shows the results of the last successful calibration.
For example…
Page 19