Cond Zero 0.00uS
@ 00/00/00 00:00
or
k=1.00 @ 2.76mS
@ 00/00/00 00:00
Note that these calibration values are still used if further measurements
are taken without re-calibrating.
7.3
Printing GLP Information to the RS232 Port
The GLP information stored in the instrument’s memory can be sent to a
printer or PC via the RS232 port.
Ensure that the
WP84
RS232 cable is connected to the instrument and to
the printer or PC.
1.
Select the GLP menu (
�
F4:Setup
�
F3:GLP
).
2.
Select
F3:Print
from the menu.
3.
The GLP information is sent to the RS232 port in formatted ASCII text.
For example…
7.4
Instrument Serial Number
In case the serial number that is fitted to the rear of the
WP-84
is
removed or becomes illegible, it is also available on the
WP-84
display.
�
The serial number is displayed at turn-on, for example…
WP84
V4.0 S1234
Con TDS Sal Temp
where
S1234
is the serial number.
�
The serial number is displayed when recalling the GLP information
(section 7.1).
�
The serial number is included on the print-out of GLP information
(section 7.3).
7.5
Additional GLP Features
Another GLP requirement is to record the date and time of every reading.
The
WP-84
does this when readings are recorded with the Notepad
function (section 8) or the Automatic Logging function (section 9).
Page 20