11.6 Data Format
Data is returned to the RS232 Port by the
WP-81
in the following format.
Please note that a “
” shown anywhere in this section denotes one
space.
LLLL
CCCCCCccc
PPPPPPppp
TTTTTTttt
dd/mm/yy
hh:mm:ss
where…
LLLL
is the Log Number. Maximum 4 characters, right justified.
The
WP-81
sends Zero for instant readings (section 11.2)
CCCCCC
is the Conductivity, TDS or Salinity Data. Maximum 6
characters, right justified.
ccc
is
the unit description, either “
uS
”, “
mS
”, “
ppM
”, “
ppK
”,
“%
“ or “psu”.
PPPPPP
is the pH or ORP/mV Data. Maximum 6 characters, right
justified.
ppp
is the unit description, sent as “
pH
” or “
mV
”
TTTTTT
is the Temperature Data. Maximum 6 characters, right
justified.
ttt
is the Temperature unit description. The
WP-81
sends
“
oC
” for real temperature data, or “
oCm
” when manual
temperature compensation is being used.
dd/mm/yy
is the date, month and year data.
hh:mm:ss
is the hours, minutes and seconds data.
Notes
1. When requested by a PC with the ?D or ?R commands (section
11.5), the data is terminated with a carriage return.
2. When the data is sent by the
WP-81
using the Print function (section
9.4) or the Instant Send function (section 11.2), the data ends with a
carriage return and a line feed.
3. When the Temperature is not within the ATC limits for
Conductivity/TDS/Salinity or pH measurement “ATCLIM” will be sent
instead of the Data.