Page 19
10. Troubleshooting
10.1 General Error Messages
Error Message
Possible Causes
Remedy
Not Factory Cal.
Not Factory Cal.
(displayed at turn-on
)
The EEPROM chip which
contains the factory
calibration information has
failed.
Switch the
901-PH
off,
wait 10 seconds, and try
switching on again.
If message persists, then the
unit must be returned to
TPS for service.
EEPROM WriteFail
EEPROM WriteFail
then:
Contact Factory
Contact Factory
(displayed at calibration
or set-up).
Storage of user calibration
settings to the EEPROM
has failed.
Switch the
901-PH
off,
wait 10 seconds, and then
switch the unit on again.
Attempt calibration/setup
again.
If message persists, then the
unit must be returned to
TPS for service.