Page 13
6. RS232 Port
This section is applicable if the optional RS232 port is fitted.
6.1 Setting the Baud Rate
1.
Select RS232 Baud Rate mode (see section 2).
2.
The currently selected baud rate is displayed. For example…
Baud Rate
Baud Rate
↑↑
9600
9600
↓↓
Press the
and
keys to scroll through the available baud rates of 300,
1200 or 9600 baud.
Ensure that the displayed baud rate matches the baud rate set on the printer
or PC with which the
901-PH
is communicating.
3.
Press the
key to return to pH, mV or Temperature mode as required.
6.2 Sending Readings to the RS232 Port
The
901-PH
can send readings to the RS232 port at a pre-set rate.
To set this Send Rate…
1.
Select RS232 Send rate mode (see section 2).
2.
The currently selected Send rate is displayed. For example…
Send Rate
Send Rate
↑↑
0
0
↓
Press the
key to increase the Send Rate.
Press the
key to decrease the Send Rate.
The Send Rate can be set from 0 to 9999 seconds.
Set the Send Rate to Zero to allow the
901-PH
to accept commands from a
remote computer.
3.
Press the