Let By Pressure Test Mode (continued)
8. After the countdown is displayed “Pass” and “Fail” will be displayed. If the
pressure bled down out of specficiation for the device under test then “Fail”
should be pressed (Down Arrow Key) and the test is over. The results can be
printed by pressing the print key or stored by pressing the save key.
9. If the device did not leak down then “Pass” is pressed (Up Arrow Key) and
the screen changes to “Tightness”. Press Go (Enter Key).
10. Once “GO” is pressed the analyzer will begin a 120 second countdown
and Wait will display in the lower left corner of the display.
11. After the countdown is displayed “Pass” and “Fail” will be displayed. If
the pressure bled down out of specficiation for the device under test then
“Fail” should be pressed (Down Arrow Key) and the test is over. The results
can be printed by pressing the print key or stored by pressing the save key.
12. If the device did not leak down then “Pass” is pressed (Up Arrow Key)
and the screen changes to “Finish...”. The test results can be printed by
pressing the print key or stored by pressing the save key.
CO Room Test
In this mode the 717 will log a Carbon Monoxide reading once per minute.
This can be used to monitor CO levels in a room over a period of time.
1. With the analyzer running, press the Func Key repeatedly to access the CO
room test function.
2. Press Start (Enter Key).
3. The 717 will begin saving CO readings once per minute for up to 50 read-
ings.
4. Press Stop (Enter Key) or wait until 50 readings have been saved.
5. Test results can be saved by pressing the Save Key or printed by
pressing the Print Key.