Instructions for Use
Toto Touch Control Unit Operating Guide
1. Starts, stops and pauses the Control Unit.
2. Air Cell side selection - Left only, Right only or Both.
3. Air Cell inflation time - Length of time the air cell is inflated; minimum 30 minutes, maximum 240
minutes.
4. Air cell rest time - Length of time the air cells are deflated and the platform is supine; minimum
30 minutes, maximum 240 minutes.
5. Mute and interface lock mode activation.
Turning the Toto Touch Control Unit On
• Press and hold the start / stop button for 3 seconds.
• A single beep will confirm activation and the start / stop green LED will be lit.
• Note the system default setting will automatically be selected.
o
Air cell inflation - Both sides
o
Air cell inflation time - 120 minutes
o
Air cell rest time - 120 minutes
• The pump will engage and the patients left air cell will start to inflate.
Air Cell Inflation Selection
• Press the Air Cell side selection button to scroll through the three available options:
o
Left and Right air cell
o
Left air cell only
o
Right air cell only
• Note: Refers to patient’s side, as stood at the foot end of the bed.
• A single beep and a lit green LED identifies the selected option.