Inspection and care of the platform and cover
Inspect the inner and outer surfaces of the cover and zip fasteners regularly for signs of damage.
Damaged covers should be reported to the ward or department manager or relevant healthcare
professional.
Torn covers should be withdrawn from service, and replacements sought.
Regularly inspect the interior and exterior of the platform, ensuring that the hinges and air cells
remain intact.
The labels on the Toto lateral turning system cover, platform and Control Unit have a unique batch
number which should be retained and used in the event of a warranty claim. NEVER remove the
labels.
Regularly check that the platform has not drifted out of position and the hoses are not blocked,
kinked or twisted.
Fitting / Replacing Damaged Air Cells
In the event of puncture or leak, it is possible to replace either or both air cells. Please contact
Frontier Therapeutics Customer Services for replacement parts Tel: +44 (0) 1495 235800.
1. Switch off the Control Unit and disconnect from mains.
2. Disconnect the air hoses to the platform using the quick connectors.
3. Unzip and pull back the cover revealing the platform.
4. Determine the faulty air cell and remove by unfastening the Velcro® straps.
5. Inspect the platform for other signs of damage, including what could have caused air cells to
have become damaged.
6. Using the Velcro® straps, fit the new air cell ensuring it is in-line with the correct section.
7. Replace the cover ensuring correct orientation and feed air cell hoses through apertures at the
bottom of the cover.
8. Reattach the hoses using the quick connectors - 'Red to Red' - ensuring they are not twisted or
kinked.
Disposal
! IMPORTANT
In order to minimise hazards to health and the environment and ensure that
materials can be recycled, this product should be disposed of at a separate collection facility for
waste electronic equipment. The Wheelie Bin symbol marked on the product is there to remind
you.
Storage Guidance
When not in use, the Toto lateral turning system should be stored in a secure location away from
the public.
It is recommended that the Toto platform is handled by a minimum of two people.
• Do not drag.
• Never store other items on top of the Toto platform.
• Place a protective covering over the platform and do not store directly on the floor.
• Do not store next to radiators or other heating devices.
• Do not allow sharp objects to penetrate the cover.
• Do not store in damp conditions.
• Remove the cover during transportation and storage.
Inspection and Care of the Toto Touch Control Unit
The Toto Touch Control Unit is designed to be reliable and long lasting, with few parts requiring
maintenance.
Air filters should be cleaned at least every 24 months and inspections should be undertaken noting
any visible signs of damage.
In the event of product failure, please contact Frontier Therapeutics for advice and recommendations
on service and repair.
Although there are no requirements for PAT testing, guideline recommendations indicate:
• Class II moveable equipment such as extension leads and portable electrical equipment should
be tested every six months. This is the responsibility of the operator alone.
Please refer to local guidance.