Care

Inspection and care of the platform and cover

Inspect the inner and outer surfaces of the cover and zip fasteners regularly for signs of damage.

Damaged covers should be reported to the ward or department manager or relevant healthcare 
professional.

Torn covers should be withdrawn from service, and replacements sought.

Regularly inspect the interior and exterior of the platform, ensuring that the hinges and air cells 
remain intact.

The labels on the Toto lateral turning system cover, platform and Control Unit have a unique batch 
number which should be retained and used in the event of a warranty claim. NEVER remove the 
labels.

Regularly check that the platform has not drifted out of position and the hoses are not blocked, 
kinked or twisted.

Fitting / Replacing Damaged Air Cells

In the event of puncture or leak, it is possible to replace either or both air cells. Please contact 
Frontier Therapeutics Customer Services for replacement parts Tel: +44 (0) 1495 235800.

1.  Switch off the Control Unit and disconnect from mains.

2.  Disconnect the air hoses to the platform using the quick connectors.

3.  Unzip and pull back the cover revealing the platform.

4.  Determine the faulty air cell and remove by unfastening the Velcro® straps.

5.  Inspect the platform for other signs of damage, including what could have caused air cells to 

have become damaged.

6.  Using the Velcro® straps, fit the new air cell ensuring it is in-line with the correct section.

7.  Replace the cover ensuring correct orientation and feed air cell hoses through apertures at the 

bottom of the cover.

8.  Reattach the hoses using the quick connectors - 'Red to Red' - ensuring they are not twisted or 

kinked.

Disposal

 !  IMPORTANT

 In order to minimise hazards to health and the environment and ensure that 

materials can be recycled, this product should be disposed of at a separate collection facility for 
waste electronic equipment. The Wheelie Bin symbol marked on the product is there to remind 
you.

Storage Guidance

When not in use, the Toto lateral turning system should be stored in a secure location away from 
the public.

It is recommended that the Toto platform is handled by a minimum of two people.

•  Do not drag.

•  Never store other items on top of the Toto platform.

•  Place a protective covering over the platform and do not store directly on the floor.

•  Do not store next to radiators or other heating devices.

•  Do not allow sharp objects to penetrate the cover.

•  Do not store in damp conditions. 

•  Remove the cover during transportation and storage.

Inspection and Care of the Toto Touch Control Unit

The Toto Touch Control Unit is designed to be reliable and long lasting, with few parts requiring 
maintenance.

Air filters should be cleaned at least every 24 months and inspections should be undertaken noting 
any visible signs of damage.

In the event of product failure, please contact Frontier Therapeutics for advice and recommendations 
on service and repair.

Although there are no requirements for PAT testing, guideline recommendations indicate:

•  Class II moveable equipment such as extension leads and portable electrical equipment should 

be tested every six months. This is the responsibility of the operator alone.

Please refer to local guidance.

Summary of Contents for GP0133-A50-01

Page 1: ...Toto Lateral Turning System with Toto Touch Control Unit User Guide www frontier group co uk...

Page 2: ...wed Do not pierce or cut Do not iron Line dry Refer to user manual CE Compliant Type BF applied part Keep dry Class II equipment Disposal do not dispose of this product as unsorted domestic waste It m...

Page 3: ...lease Pull Tag 45 XL 35 1 4 271 11 Quick Release Pull Tag 30 Product Weight kg lbs Maximum User Weight kg lbs Width mm Inches Length mm Inches Standard 18 40 190 419 745 29 1850 73 45 18 40 190 419 74...

Page 4: ...es or pillows etc Intended Care Setting Toto Lateral turning system is intended to be used in the following care settings Long term or extended care such as nursing homes and care homes Home care Hosp...

Page 5: ...the Control Unit Black to Control Unit ensuring the hoses are not twisted or kinked F 10 Connect the Control Unit to the mains power via the cable supplied G A The platform should be placed directly o...

Page 6: ...d hold the start stop button for 3 seconds A single beep will confirm activation and the start stop green LED will be lit Note the system default setting will automatically be selected o Air cell infl...

Page 7: ...utton a single beep will confirm activation The start stop LED will flash when active and all other LED s will turn off After 10 minutes in Therapy Pause Mode the alarm will sound To continue in this...

Page 8: ...dible beep and Low system pressure Pressure in the system has fallen below the minimum operational requirements Five Rest time LEDs flashing when activated High system pressure Air flow between the Co...

Page 9: ...ion even though the service LED remains lit Interface panel is lit but unresponsive Interface panel is locked Press and hold ALARM MUTE button for 4 seconds Audible beep no LEDs lit Accidental removal...

Page 10: ...on and feed air cell hoses through apertures at the bottom of the cover 8 Reattach the hoses using the quick connectors Red to Red ensuring they are not twisted or kinked Disposal IMPORTANT In order t...

Page 11: ...n may be achieved by laundering at temperatures not exceeding 65 C for 10 minutes or 73 C for three minutes 1 1 Protect Rinse and Dry BHTA guidance on the care cleaning and inspection of healthcare ma...

Page 12: ...ot affect your statutory rights Fire Testing The Toto lateral turning system platform meets the requirements of The Furniture Furnishings Fire Safety Regulations 1988 Meeting the requirements of BS 71...

Page 13: ...rk of Frontier Therapeutics Ltd November 2017 Frontier Therapeutics Limited All rights reserved GP0133_A50_01 w ww your turn org u k Frontier Therapeutics Limited Newbridge Road Industrial Estate Blac...

Reviews: