Cleaning

Toto Lateral Turning System Platform Cover

Keeping a cleaning record is recommended as the product is suitable for re-use.

Light Soiling

 !  IMPORTANT

 

In the absence of gross contamination or unusual risk, the removal of dirt and 

spillages with a neutral detergent and warm water should be sufficient. A 0.1% hypochlorite solution 
(1,000ppm) may also be used. Ensure that the cover is rinsed with clean water using a single use, 
non-abrasive cloth and thoroughly dried. Ensure adequate ventilation.

Heavy Soiling

 !  IMPORTANT

 In cases of gross contamination, the Toto lateral turning system platform cover 

should be cleaned first with detergent and water and then with a 0.1% hypochlorite solution 
(1,000ppm). Rinse well with clean water and wipe down with a single use, non-abrasive cloth and 
thoroughly dry. Ensure adequate ventilation.

 !   IMPORTANT

 If required, a 1% hypochlorite solution (10,000ppm) and cold water may be applied 

for no more than two minutes.

Machine Washing Instructions

  !   IMPORTANT

 Disinfection may be achieved by laundering at temperatures not exceeding 65ºC 

for 10 minutes or 73ºC for three minutes.

1

1

Protect, Rinse and Dry - BHTA guidance on the care, cleaning and inspection of healthcare 

mattresses. (January 2012 - BHTA HM 0112)

Drying

To avoid shrinkage, line dry in a clean, indoor environment. Covers must be thoroughly dried 
before re-fitting to the Toto platform.

Do not mangle or iron.

Platform, Cabling, Control Unit and Air Hoses

Before cleaning, disconnect from mains power supply.

The platform, cabling, Control Unit and air hoses can be wiped clean with alcohol wipes and/or a 
chlorine derivative. Care should be taken when wiping the Control Unit not to allow liquids to enter 
through any apertures.

Cleaning Air Filters

1.  Switch off the power supply to the Control Unit.

2.  Disconnect the power lead and air hoses.

3.  Place the Control Unit on a flat surface with its back panel uppermost (place a soft cloth under 

the unit to prevent scratches).

4.  Carefully remove the air filter cover located on the back panel of the Control Unit (x1).

5.  Remove and clean the filter with clean water and allow to fully dry.

6.  Refit the air filter cover to the Control Unit.

7.  The Control Unit is now ready for re-use.

Summary of Contents for GP0133-A50-01

Page 1: ...Toto Lateral Turning System with Toto Touch Control Unit User Guide www frontier group co uk...

Page 2: ...wed Do not pierce or cut Do not iron Line dry Refer to user manual CE Compliant Type BF applied part Keep dry Class II equipment Disposal do not dispose of this product as unsorted domestic waste It m...

Page 3: ...lease Pull Tag 45 XL 35 1 4 271 11 Quick Release Pull Tag 30 Product Weight kg lbs Maximum User Weight kg lbs Width mm Inches Length mm Inches Standard 18 40 190 419 745 29 1850 73 45 18 40 190 419 74...

Page 4: ...es or pillows etc Intended Care Setting Toto Lateral turning system is intended to be used in the following care settings Long term or extended care such as nursing homes and care homes Home care Hosp...

Page 5: ...the Control Unit Black to Control Unit ensuring the hoses are not twisted or kinked F 10 Connect the Control Unit to the mains power via the cable supplied G A The platform should be placed directly o...

Page 6: ...d hold the start stop button for 3 seconds A single beep will confirm activation and the start stop green LED will be lit Note the system default setting will automatically be selected o Air cell infl...

Page 7: ...utton a single beep will confirm activation The start stop LED will flash when active and all other LED s will turn off After 10 minutes in Therapy Pause Mode the alarm will sound To continue in this...

Page 8: ...dible beep and Low system pressure Pressure in the system has fallen below the minimum operational requirements Five Rest time LEDs flashing when activated High system pressure Air flow between the Co...

Page 9: ...ion even though the service LED remains lit Interface panel is lit but unresponsive Interface panel is locked Press and hold ALARM MUTE button for 4 seconds Audible beep no LEDs lit Accidental removal...

Page 10: ...on and feed air cell hoses through apertures at the bottom of the cover 8 Reattach the hoses using the quick connectors Red to Red ensuring they are not twisted or kinked Disposal IMPORTANT In order t...

Page 11: ...n may be achieved by laundering at temperatures not exceeding 65 C for 10 minutes or 73 C for three minutes 1 1 Protect Rinse and Dry BHTA guidance on the care cleaning and inspection of healthcare ma...

Page 12: ...ot affect your statutory rights Fire Testing The Toto lateral turning system platform meets the requirements of The Furniture Furnishings Fire Safety Regulations 1988 Meeting the requirements of BS 71...

Page 13: ...rk of Frontier Therapeutics Ltd November 2017 Frontier Therapeutics Limited All rights reserved GP0133_A50_01 w ww your turn org u k Frontier Therapeutics Limited Newbridge Road Industrial Estate Blac...

Reviews: