TOTAL Lubmarine Tech'Care User Manual Download Page 10

The Rapid Oil Test Gauge has an internal clock which is used to time stamp test results when they are stored in the instrument memory.

To adjust the internal clock:

8

Press any of the 4 front panel keys to 

switch the Rapid Optical Oil Tester on.
The instrument initializes and displays its

serial number and embedded software

version for a few seconds.
The instrument then displays the Main

Run Screen which offers three functions:

[Conf]

,

 [Test]

 and

 [Hist]

5

 - The year is displayed.

Press the [

] key to increment the year.

Press the [

] key to decrement the year.

Press [Next] when done.

- Press any of the 4 front panel keys to 

turn the Rapid Optical Oil Tester on.

- Press the [Conf] key.

- Press the [Pwr Off] key.

Press the key corresponding to the

function you want to run:

Select 

[Conf]

 to adjust the internal clock, 

the oil monitoring function and the power 

off time of your instrument.

Select 

[Test]

 to run a water test or a BN

test (optional)

Select 

[Hist]

 to view the test history.

Refer to the appropriate section of this

manual for details on each function.

In order to preserve battery life the instru-

ment turns off automatically if not used for 

a period of time. Refer to section 2 below 

for instructions for setting power off time.

The user may also press the      key for 

2 seconds at any time to switch off the 

Rapid Optical Oil Tester
Note: When first powered on after the 

battery has been replaced, the instrument 

accesses automatically the clock adjust-

ment function.

1

 - Press any of the 4 front panel keys to

turn the Rapid Optical Oil Tester on.

2

 - Press the [Conf] key.

3

 - Press the [Time] key.

Note:

When first powered on after the battery has been replaced, 

the instrument accesses automatically the clock adjustment function.

6

 - Adjust the month and press [Next]

when done.

- Adjust the day and press [Next] when done.

8

 - Adjust the hour and press [Next]

when done.

9

 - Adjust the minutes and press [Save]

to store the data and return to the main

run screen.

At any time press briefly the       key to 

exit function and return to the Main Run 

Screen. Modifications will not be saved.

The Rapid Optical Oil Tester shows the current Power Off time.

- Press the [

] key to increase the value.

Press the [

] key to decrease the value.

2 - 

Press [Save] when done

At any time press briefly the      key to exit function and return 

to the Main Run Screen. Modifications will not be saved.

 

Alarm

Pwr Off

Time

Config

Next

Year : 2019

Set Date

Save

Minute : 30

Set Time 

 

Alarm

Pwr Off

Time

Config

Save

2 minute

Set Power Off

LUBMARINE

Test

Conf

Hist

RAPID OPTICAL

OIL TESTER

 

Rapid Optical Oil Tester

Operating the Rapid Optical Oil tester

3 Adjusting the internal clock

1 Power on / Main run screen / Power off

2 Adjusting power off time

Summary of Contents for Tech'Care

Page 1: ......

Page 2: ......

Page 3: ...e test history to a computer 5 3 Erasing the test history 100 l precision pipette Iron Test Safety data sheets Replenishment of consumables Pages 4 5 6 7 8 9 10 11 13 19 21 Overview The Lubmarine Tech...

Page 4: ......

Page 5: ...os User Guide Pen Syringes Magnetic rack Wipe felt pad 100 l precision pipette Pipetting level adjuster Rapid Optical Oil tester Mini USB cable Vial caps Reagent IT1 Solvent IT2 Vials Solution IT3 Spa...

Page 6: ...sing a front panel display and special function keys A USB communication port can be used to download test results to a computer 4 1 Product overview 2 Sub Assemblies description and location Keyboard...

Page 7: ...test evaluation date and time Lubmarine reagents IT1 IT2 IT3 100 l 3 drops 60 to 800 ppm iron 1 ppm Data valid for a 100 l oil sample 3 drops only 5 Declaration of Conformity We GESERCO SARL 16 RUE F...

Page 8: ...6LR6I PP3 battery which powers your Rapid Oil Test Gauge should provide two years operation or more LUBMARINE Test Conf Hist RAPID OPTICAL OIL TESTER Rapid Optical Oil Tester Product Overwiew A batter...

Page 9: ...preserve the instrument over time the Rapid Optical Oil Tester must always be kept clean and dry The test port cap should be replaced when the Rapid Optical Oil Tester is not being used in order to p...

Page 10: ...also press the key for 2 seconds at any time to switch off the Rapid Optical Oil Tester Note When first powered on after the battery has been replaced the instrument accesses automatically the clock...

Page 11: ...status of the function 4 Press the On Off key 5 Confirm the desired status by pressing the Save key At any time press briefly the key to exit function and return to the Main Run Screen Modifications...

Page 12: ...on can store up to 1 000 test results When the instrument memory is full a History full message shows and the instrument will not store any additional data unless the memory is cleared 3 The instrumen...

Page 13: ...e pipetting level adjuster comes to a stop Push the pipetting knob down to its stop Dip the capillary into the oil until the pipetting level adjuster comes to a stop Slowly release the pipette plunger...

Page 14: ...allow to dry completely To fit a new capillary Slide the capillary to lock position A Screw the collet by hand into place Checking for tightness Daily Draw in some oil Hold the pipette upright for 10...

Page 15: ...within 2 hours after the cylinder oil samples have been drawn Wear goggles Wear protective gloves Read the instructions and MSDS contained in this manual Before each test look after your own safety a...

Page 16: ...y Follow the test instructions displayed by the instrument Press the key for more instructions Close the vial with a new cap and flip it over once smoothly to mix the contents After mixing it is norma...

Page 17: ...t the pipetting level adjuster is mounted on to the precision pipette and locked Return the vial to the magnetic rack and remove the cap Inject oil into the vial Make sure the oil drops go into the re...

Page 18: ...the function settings Dispose of the vial and cap in an appropriate waste container Do not use them for another test as results may be biased Do not dispose of the residues in the drainage system Care...

Page 19: ...Solution IT1 Solvent IT2 Reagent IT3 17 Health Safety Environmental Data...

Page 20: ...toxication remove the person affected from the expo sure area and provide with fresh air Seek medical attention if the symptoms get worse or persist Eyes Rinse eyes thoroughly with luke warm water for...

Page 21: ...y Environmental Take all necessary precautions against accidental spillage in to soil or water PROTECTIVE CLOTHING Protective clothing including impervious gloves should be worn if skin contact is ant...

Page 22: ...on remove the person affected from the exposure area and provide with fresh air Seek medical attention if the symptoms get worse or persist Eyes Rinse eyes thoroughly with luke warm water for at least...

Page 23: ...unit 1 unit 1 kit 1 Flask 1 unit 1 complete set for 50 iron tests Code IT9111 DP6151 DP6152 IT9100 IT9107 DP6133 IT9110 PT9103 GESERCO SARL Attn Sales department 16 rue Fran ois Arago 33700 M rignac...

Page 24: ......

Reviews: