
−
58
−
6 F 8 A 0 9 3 4
8.2.4 Exciting Frequency
Proceed as follows to check the Exciting Frequency
Switch operation
Display example
Description
A:DETECTOR
1 EXC CUR
2 SIZE
3 EXC FREQ
4 FLOW DIR
5 EXIT
Select "
EXC FREQ
" from the setting item
selection menu.
STEP1
EXC FREQ
200
Hz
The currently set excitation frequency
(
200Hz
in this example) appears.
Then push .
* Pushing returns you to the setting
menu.
8.2.5 Flow Direction Setting
In the converter, you can set the flow direction of fluid arbitrarily.
●
Flow direction setting
Selection item
Contents
NORMAL
When the fluid flows in the direction of the arrow
indicating the flow direction that is attached to the
detector, the indicator value and electric current
output value increase.
SWITCH
When the fluid flows in the reverse direction of the
arrow indicating the flow direction that is attached to
the detector, the indicator value and electric current
output value increases.
Switch operation
Display example
Description
A:DETECTOR
1 EXC CUR
2 SIZE
3 EXC FREQ
4 FLOW DIR
5 EXIT
Select "
FLOW DIR
" from the setting item
selection menu.
ESC
ESC
ESC
ESC