- 41 -
6F8A0883
9. Specifications
The flowmeter specifications and the type specification code used when ordering the flowmeter
are described in this chapter.
9.1 Specifications
Meter size:
1/2", 1", 1 1/4", 1 1/2", 2", 2 1/2", 3", 4", 5", 6", 8", 10", 12", 14", 16", 18", 20" and 24"
(15, 25, 32, 40, 50, 65, 80, 100, 125, 150, 200, 250, 300, 350, 400, 450, 500 and 600 mm)
Measuring range:
0
–1.0
ft/s to 0
–
32.8
ft/s (0
–
0.3
m/s to 0
–10
m/s).
0
–
0.3
ft/s to 0
–
1.0
ft/s (0
–
0.1
m/s to 0
–
0.3
m/s) range is available optionally for meter
size 1/2” to 18” (15 to 450 mm).
Accuracy:
(Accuracy when combined with the converter)
< 1/2” to 18”( 15mm to 450 mm ) >
±0.2 % of Rate *
1
*
1
This pulse output error result is established under standard operating conditions at Toshiba's flow
calibration facility, Fuchu Japan. (NIST Traceable).
*
1
Individual meter measurement error may vary up to ±0.5% of Rate at 1.64 ft/s (0.5m/s) or more and
±0.3% of rate ±0.039 inch/s (1mm/s) at 1.64 ft/s or less.
*
1
Current output: plus ± 8µA (0.05% of span).
*
1
Refer to individual calibration data for each individual meter's measurement error.
< 20” to 24” ( 500mm to 600mm ) >
± 0.3 % of Rate
*
2
*
2
This pulse output error result is established under standard operating conditions at Toshiba’s flow
calibration facility, Fuchu Japan.
*
2
Individual meter measurement error may vary up to ±0.5 % of Rate at 3.28 ft/s (1.0 m/s) or more
and ± 0.3% of Rate ±0.079 inch/s (2 mm/s) at 3.28 ft/s (1.0 m/s) or less.
*
2
Current output: plus ± 8µA (0.05% of span.)
*
2
Refer to individual calibration data for each individual meter’s measurement error.
Fluid conductivity:
5 µS/cm minimum
Fluid temperature:
–20
゚
C to +100
゚
C (–4
゚
F to 212
゚
F)
・・・
FEP
–20
゚
C to +120
゚
C (–4
゚
F to 248
゚
F)
・・・
PTFE
–20
゚
C to +60
゚
C (–4
゚
F to 140
゚
F)
・・・
Polyurethane (PU)
–10
゚
C to +60
゚
C (14
゚
F to 140
゚
F)
・・・
Chloroprene rubber(CR)
Ambient temperature:
–20
゚
C to + 60
゚
C (–4
゚
F to 140
゚
F)
Storage temperature:
–25
゚
C to +65
゚
C (–13
゚
F to 149
゚
F)
Storage humidity:
10 to 90%RH (no condensation)
Fluid pressure:
0 to 1MPa (0 to 150 psi, or 0 to 10 bar)
(To be within the applicable flange limitation)