Guidance and manufacturer’s declaration – electromagnetic immunity
The FPI is intended for use in the electromagnetic specified below.
The customer or the user of the FPI should assure that it is used in such an environment.
Immunity
test
IEC60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC61000-4-2
±
6kV contact
±
8kV air
±
6kV contact
±
8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material.
The relative humidity should be at
least 30 %.
Electrical fast
transient
/
burst
IEC61000-4-4
±
2 kV for power
supply lines
±
1kV for input/
output lines
±
2kV for power
supply lines
±
1kV for input/
output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC61000-4-5
±
1kV differential
mode
±
2kV common
mode
±
1kV differential
mode
±
2kV common
mode
Mains power quality should be that
of a typical hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
<
5 % U
T
(
>
95 % dip in, U
T
)
for 5 sec
<
5 % U
T
(
>
95 % dip in, U
T
)
for 5 sec
Mains power quality should be that
of a typical hospital environment. If
the user of the FPI requires
continued operation during power
mains interruptions, it is
recommended that the FPI be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60Hz)
magnetic field
IEC61000-4-8
3A/m 3A/m Power
frequency magnetic fields
should be at levels characteristic
of a typical location in atypical
commercial or hospital
environment.
Note:
(1) U
T
is the a.c.mains voltage prior to application of the test level.
(2) This FPI can only allow to use the designated cables and components.
If the different cables or components are used, they may deteriorate the performance of
recommended separation distance between portable and mobile RF communications
equipment and this system.