No. 2A201-519EN*A
46
Table 8-1 List of inter-unit cables (3/3)
Signal cables (2/2)
Cable
No.
Wiring
pit
Connected
to: (1)
Connected
to: (2)
Standard
length (m)
Effective
length (m)
Retract length
(m)
Maximum
length (m)
Number of cores
×
section area (mm
2
)
Outer cable
diameter (mm)
Connector diameter
(mm)
Remarks
S05 S Console
REC BOX
Gantry 14
11
0.5
while
connecting to (1)
1.5 while
connecting to (2)
19 2-pairs,
optical
fiber
11
Connector (22
×
8
×
25) Minimum
permissible
bending
radius:
110 mm
S06 S Console
REC BOX
Gantry 14
11
0.5
while
connecting to (1)
1.5 while
connecting to (2)
19 2-pairs,
optical
fiber
11
Connector (22
×
8
×
25) Minimum
permissible
bending
radius:
110 mm
Cables attached to the gantry
Cable
No.
Wiring
pit
Connected
to: (1)
Connected
to: (2)
Standard
length (m)
Effective
length (m)
Retract length
(m)
Maximum
length (m)
Number of cores
×
section area (mm
2
)
Outer cable
diameter (mm) Connector diameter (mm)
Remarks
FCS12 S Gantry Patient
couch
⎯
⎯
⎯
⎯
25TP
×
0.08
shielding
8.1
Connector (73
×
19
×
47)
while connecting to (2)
(1)
connected
inside gantry
FCP40 P Gantry Patient
couch
⎯
⎯
⎯
⎯
2
cores
×
0.75
3 cores
×
2.0
4 cores
×
0.5
6.1
6.2
9.9
Connector (28
×
33
×
28)
Compound cable
(1)
connected
inside gantry
Compound
cable
FCE01 E Gantry Patient
couch
⎯
⎯
⎯
⎯
1
core
×
8.4
7.8 Solderless
contact
while
connecting to (2)
(1)
connected
inside gantry
FCS46 S Gantry Foot-SW
⎯
⎯
⎯
⎯
5TP
×
0.08
shielding
7.2
Connector (
φ
29
×
52) while
connecting to (2)
(1)
connected
inside gantry