VIII. SIMPLE OPERATION
The IN35 and IN45 are fully programmable heating/chilling incubators. With them the
user has the ability to write and store in memory three programs of as many as ten
steps each for instant recall and use later. Also, each incubator can be used to set a
single temperature, temperature ramp, and timed function without going into the
memory of the unit. The memory should be used to store complex routines where the
temperature can be changed automatically to other temperatures, at given rates of
change if wanted, and held at those temperatures for specific times.
IX. SIMPLE TEMPERATURE, TEMPERATURE RAMP, TIMER, AND AUTO-OFF
SETTINGS
All of these functions are easily set from the keyboard. Please note that the keyboard
has both tactile and audible feedback.
SETTING TEMPERATURE
1.
Depress and release the TEMP key. The display will show three zeros to the left
of the actual chamber temperature. The digit on the right of the three will be
flashing.
2.
Use the LEFT ARROW in the center of the membrane switch to move the
settable digit to the left. Now use the UP ARROW or DOWN ARROW to scroll to
the temperature setting wanted for that digit. Use the RIGHT ARROW to move
to the next digit to the right. Use the UP ARROW or DOWN ARROW to set that
digit. Use the RIGHT ARROW to move to the right to set the final digit if wanted.
3
Depress ENTER. The target temperature is now set. The actual chamber
temperature as displayed in the center of the top row of the display will now start
to change and go toward the target.
4
To remove the temperature target hit TEMP, ERASE. The target temperature is
now gone and the unit will stop heating or chilling as previously instructed.
Note: If you wish to cancel an instruction made before entering the value,
hit CANCEL. To achieve the minimum temperature of 4ºC the ambient
temperature of your laboratory should not be over 24ºC. The unit has a
below ambient temperature range from ambient room temperature. This
range is about 18ºC to 20ºC and is load dependent. It may take longer to
reach very low temperatures when
the chamber is fully loaded.
The unit has a power failure backup circuit. When power fails and then
returns, the unit will continue to run at the temperature it was running when
power failed.