4. 

Position the unit on a bench where it has at least 6" (15.24 cm) clearance from 

 

the back of the unit to the wall, bench or other obstruction.  The cooling fan on 

 

the rear of the unit needs to draw air clearly to cool the Peltier unit.  Also, there 

 

must be about 3" (7.62 cm) clearance on either side of the unit and 6” (15.24 cm) 

 

on top of the unit to allow proper venting.  

 
5. 

Place the wire racks provided into the slots selected.   

 
6. 

Turn the unit on by the main On/Off switch on the rear panel.  The LCD on the 
front panel will illuminate and be in the temperature/timer display mode.  The 
fans on the rear panel and inside the instrument chamber will start to run.  The 
instrument is now ready to use. 

 

VI. GENERAL DESCRIPTION 

 
The Torrey Pines Scientific Chilling Incubators are digitally controlled and Peltier driven.  
Because they are Peltier driven, they have no CFC’s or other refrigerating gases.  Also, 
being Peltier driven, the units have the ability to heat and cool

 and control exactly at 

or near ambient room temperature.  

The units can hold temperatures accurately, and 

reproduce temperatures exactly.  The chamber is molded ABS plastic and the racks are 
electro-polished steel.   
 

 

 

 

 

 

 

Since the units are Peltier driven there is a fan in the chamber and one on the rear of 
the units.  The internal fan is used to keep the chamber at a uniform temperature 
throughout.  There is an inner chamber door of clear lexan to help hold the chamber 
temperature while viewing the sample. 
 
Temperature is sensed by a microchip inside the chamber.  The settable temperature 
range is from 4.0ºC to 70.0ºC.  The unit has a membrane keypad with tactile touch and 
audible feedback for setting temperature, temperature ramp, timer, auto-off, and three 
programs of ten steps each.   
 
The timer is in days, hours, minutes and seconds. The timer can be set to 99:59:59:59 
days, hours, minutes, and seconds maximum.  When the timer counts to zero it will 
sound the audible alarm for one minute and the timer will start to count up. The unit can 
be set to turn off the heater/chiller at the end of a timed event by using the AUTO-OFF 
function.   
 
The IN35 and IN45 are supplied with an RS232 port for recording temperature data at 
one data point per second.  More information is available on this in a later chapter. 
 

 
 
 
 

Summary of Contents for EchoTherm IN35-12VDC

Page 1: ...TOR MODELS IN35 AND IN45 12VDC DOCUMENT NUMBER IN35 03 Revised 15 February 2012 TORREY PINES SCIENTIFIC INC 2713 Loker Ave West Carlsbad CA 92010 Telephone 760 930 9400 Fax 760 930 9480 E Mail info to...

Page 2: ...F ITEMS 3 IV CAUTIONS 3 4 V SET UP 4 5 VI GENERAL DESCRIPTION 5 OPERATION VII FRONT AND REAR PANEL 6 7 VIII SIMPLE OPERATION 8 IX SIMPLE TEMP TIMER RAMP AND AUTO OFF SETTINGS 8 9 10 X PROGRAMMABLE OPE...

Page 3: ...eturned For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss Torrey Pines Scientific will not be liable for damage resulting fro...

Page 4: ...n toward the front of the chamber where it can easily be cleared GENERAL The IN35 and IN45 are designed for laboratory use and for use only as instructed in this manual If the instrument is used in a...

Page 5: ...hold temperatures accurately and reproduce temperatures exactly The chamber is molded ABS plastic and the racks are electro polished steel Since the units are Peltier driven there is a fan in the cham...

Page 6: ...ture in the center of the top line of the display and the timer with all zeros across the bottom line The display is a two line 16 digit alphanumeric back lighted LCD When a target temperature has bee...

Page 7: ...RS232 and a reorder label The I O port will be discussed later in the chapter on the RS232 The label tells the model number serial number operating voltage of the unit power consumption and the addres...

Page 8: ...ove the settable digit to the left Now use the UP ARROW or DOWN ARROW to scroll to the temperature setting wanted for that digit Use the RIGHT ARROW to move to the next digit to the right Use the UP A...

Page 9: ...or RIGHT ARROW 3 When the value is set press ENTER The ramp rate is now set SETTING TIMER The timer is a count down timer that can be set to a maximum of 99 days 59 hours 59 minutes and 59 seconds It...

Page 10: ...R will erase the programs and the display will show PLEASE WAIT The programs will then be erased This erases ALL programs in memory Hitting CANCEL will cancel the instruction To change or erase one pr...

Page 11: ...m hit the UP ARROW 7 After the last program step data has been entered press ENTER The display will now show CYCLE 01 This gives the user the choice of having the program just entered cycle one or mor...

Page 12: ...NCEL on the bottom line 3 Depress ENTER 4 Enter the value onto the display as read from your temperature standard and depress ENTER Single point calibration is now complete and will be held in memory...

Page 13: ...the RS232 interface capable of doing this The instructions for connection and use follow RS232 INTERFACE SPECIFICATION Parameters 2400 baud 8 data bits 1 stop bit no parity No handshake hardware or s...

Page 14: ......

Reviews: