INSTALLATION AND MODE OF OPERATION
3
INSTALLATION AND MODE OF OPERATION
3.1 Installation notes
Information!
All instruments are carefully checked for proper function before shipment.
Check immediately on receipt, the outer packing carefully for damage or
signs of improper handling.
Report damage to the carrier and your locale sale staff. In such cases, a
description of the defect, the type and the serial number of the device is
indicated.
Unpack the unit carefully to avoid damage.
Check the completeness of the delivery against the packing list. Check the
name plate, if the delivered flow meter according to your order.
3.2
Installation instructions
Installation of the orifice plate
Before and after the orifice plate a straight calming section is provided as a function of
the nominal diameter (D).
With an installation of an one quarter pipe (independend of the direction) the distance of
the inlet path of min. 5 x D and an outflow zone of 2 x D is required. The connected pipes
have to be of the same size as the orifice plate.