APPENDIX C. Parameters to be set during on-site con-
figuration
Table C.1
Representation
of the parame-
ter
on PC screen
Parameter Name,
measurement units
Possible
values
Value
after initializa-
tion
Notes
1
2
3
4
5
Menu
"System"
Address
Flowmeter address in the RS-
interface network
1-247
1
Delay, ms
Response lag in the RS-interface
network
0-125
50
Time and Date
Current flowmeter time and date
XX/XX/XX
(dd.mm.yy)
XX:XX:XX
(hh:mm:ss)
No changes
available
Mode
Clock transition between "Sum-
mer" and "Winter" time
no switch;
standard;
user
no switch
Data log clear
Clearing logs
...; start
...
Total flow
reset
Zero setting for volume totalizer
...; start
...
Menu
"Service"
Median mean
Number of measurements to de-
termine the median value
1-21
7
Arithmetic
mean
Number of measurements to de-
termine the arithmetic mean val-
ue
1-50
25
Inertion time, s
Minimal duration of the event re-
corded to logs
0-255
0
Cutoff
flowrate, m
3
/h
Minimal flowrate cutoff
0,000-999,000
flowmeter sen-
sitivity threshold
Battery control
Battery status monitoring
off;
3,2; 3,35; 3,5
off
Contrast of
the LCD
LCD contrast setting
1-63
28
Work time of
LCD, s
Time to turn off the LCD after the
last pressing of a button
1-255
30
Flowrate units
Flowrate measurement units
m
3
/h;
l/min
m
3
/h
35