![Topward 6302A, 6302AR, 6302D, 6302DR, 6303A, 6303AR, 6303D,6303DR, 6306A, 6306AR, 6306D, 6306DR, 6603A, 6603AR, 6603D, and 6603D Operation Manual Download Page 14](http://html1.mh-extra.com/html/topward/6302a-6302ar-6302d-6302dr-6303a-6303ar-6303d-6303dr-6306a-6306ar-6306d-6306dr-6603a-6603ar-6603d-and-6603d/6302a-6302ar-6302d-6302dr-6303a-6303ar-6303d-6303dr-6306a-6306ar-6306d-6306dr-6603a-6603ar-6603d-and-6603d_operation-manual_1129994014.webp)
6000 Series
3.6 Serial
Operation
Serial connection is used to connect two or more power supplies in series to obtain a higher voltage ratings
(maximum 240V for this power supply).
Figure 7 shows the serial connection between two 6303D power supplies at series tracking. One unit is set at
60V/2A (MASTER setting is 30V/2A) while the other is at 40V/3A (MASTER setting is 20V/3A). Figure 8
shows the serial connection voltage/current output versus load changes.
+
-
V
A
+
-
+
-
5V5A
OVER LOAD
V1=VS1+VM1
VO = V1 + V2
+
-
V
A
+
-
+
-
5V5A
OVER LOAD
V2=VS2+VM
Serial Tracking Mode
Serial Tracking Mode
SET : V1 = 60V / 2A
V
20
40
60
80
100
60
50
40
30
20 13.3
OHM
V1
V2
TOTAL OUTPUT VOLTAGE(VO)
V1 C.V. MODE
V1 C.C. MODE
V2 C.V. MODE
V2 C.C. MODE
SET : V2 = 40V / 3A
60
50
40
30
20 13.3
OHM
1A
2A
3A
Figure 7 Serial Connection Diagram
Figure 8 Serial Voltage/Current Output vs. Load
Change
11