Technical details electrical components
Manufacturer: Concens AS, Denmark
Controller box: C3con1-x
Actuator: Con 5
Battery: C3 NiMH 24 V – 1400 mAh
Battery charger: C3, Input voltage: 100–240VAC
Input frequency: 50/60Hz
Input current: Max 200mA (fuse 1A)
Output current: Max 400mA
Working ambient temperature for actuator: −5 °C to +40 °C
Working ambient temperature for controller and battery
charger: 0 °C to +40 °C
Humidity: 20–80 %
Delivery and storing ambient temperature for actuator,
controller and battery charger: −20 °C to +50 °C
Humidity: 20–80 %
Guidance and manufacturer’s declaration
Electromagnetic EMISSIONS for all ME EQUIPMENT and ME SYSTEMS
The TOPRO TAURUS walking aid is intended for use in the electromagnetic environment specified below.
The customer or the user of the TOPRO TAURUS walking aid should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The TOPRO TAURUS walking aid uses RF energy only for
its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The TOPRO TAURUS walking aid is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration
Electromagnetic IMMUNITY for all ME EQUIPMENT and ME SYSTEMS
The TOPRO TAURUS is intended for use in the electromagnetic environment specified below.
The customer or the user of the TOPRO TAURUS should assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relati-
ve humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
Main power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
Main power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Main power quality should be that of a typical
commercial or hospital environment. If the user of
the TOPRO TAURUS requires continued operation
during power mains interruptions, it is recommen-
ded that the TOPRO TAURUS be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Power frequency magnetic fields should be at le-
vels characteristic of a typical location in a typical
commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.