FUNCTION: RESISTANCE MEASUREMENT
RANGE
RESOLUTION
ACCURACY
200Ω
100mΩ
+- (1,0% + 2)
2000Ω
1Ω
+- (0,8% + 2)
20kΩ
10Ω
+- (0,8% + 2)
200kΩ
100Ω
+- (0,8% + 2)
2000kΩ
1kΩ
+- (1,2% + 2)
FUNCTION: SEMICONDUCTOR DIODE TEST
RANGE
DESCRIPTION
LCD display shows diode forward voltage.
FUNCTION: BATTERY TEST
RANGE
DESCRIPTION
TEST CONDITIONS
1.5V
Display shows battery voltage. It allows to
estimate battery level.
Operating current approx. 20mA
9V
1.5V or 9V.
Operating current approx. 5mA
TAKING MEASUREMENTS
DIRECT VOLTAGE MEASUREMENT
1. Connect red test lead to VmAΩ terminal, black test lead to COM terminal.
2. Set the dial function selector to DC voltage „
„ and correct range corresponding to
measured value. If you do not know range of DC voltage set the selector switch to maxi-
mum range and then reduce it gradually for the best measurement accuracy.
3. Connect or press test probes to measurement spots in the device or circuit. Measured volt-
age value and its polarity will be displayed on the LCD display.
AC VOLTAGE RMS MEASUREMENT
1. Connect red test lead to VmAΩ terminal, black test lead to COM terminal.
2. Set the dial function selector to AC voltage „
„ and correct range corresponding to
measured value. If you do not know range of AC voltage set the selector switch to maxi-
mum range and then reduce it gradually for the best measurement accuracy.
3. Connect or press test probes to measurement spots in the device or circuit. Measured volt-
age value and its polarity will be shown on the LCD display.
DIRECT CURRENT MEASUREMENT
1. Connect red test lead to VmAΩ terminal, and black test lead to COM terminal if measured
current is not greater than 200mA. If you expect the measured current to exceed 200mA
connect red test lead to “10A” terminal. Maximum measurement time with 10A current
flow is 10 seconds. Wait for at least 15 minutes before taking next current measurement
with the use of 10A range.
Summary of Contents for 94W101
Page 18: ...18 TOPEX PL 94W101 94W101 IEC 1010 II CAT II 600 II 200 AC DC II RU...
Page 19: ...19 TOPEX PL 1 2 3 4 100 5 6 7 8 9 10 11 12 13 14 40 20 rms...
Page 20: ...20 TOPEX PL 15 16 17 600 10A 18 II III IV 19 1 2 3 5 4 1 3 1999 2 OFF OFF...
Page 23: ...23 TOPEX PL 3 1 VmA COM 2 3 1 VmA COM 200 10A 10A 10 15 10A 2 3 4 1 VmA COM 2...
Page 24: ...24 TOPEX PL 3 4 1 VmA COM 2 3 4 5 6 1 VmA COM 2 1 5 9 3 1 2 OFF 3 2 4 6F22 1 OFF 2 2 F250 250...
Page 25: ...25 TOPEX PL...
Page 26: ...26 TOPEX PL 94W101 94W101 IEC 1010 II CAT II 600V II 200 AC DC II UA...
Page 27: ...27 TOPEX PL 1 2 3 4 5 6 7 8 9 10 11...
Page 28: ...28 TOPEX PL 12 13 14 40 20 rms 15 16 17 600 10A 18 II III IV 19 1 2 3 5 4...
Page 32: ...32 TOPEX PL 2 3 4 1 VmA COM 2 3 4 1 VmA COM 2 3 4 5 6 1 VmA COM 2 1 5 9...
Page 33: ...33 TOPEX PL 3 1 2 OFF 3 2 4 6F22 1 OFF 2 2 F250 250...
Page 71: ...71 TOPEX PL...