46
BASIC OPERATION
6
While keeping the alignment dot within the
alignment frame, push the instrument
toward the patient a little bit at a time.
When the instrument approaches the
anterior eye segment to some extent, the
image switches to the endothelium obser-
vation display and the longitudinal align-
ment bar appears on the screen.
The longitudinal alignment bar is displayed with a dotted line when the instrument is close to the
anterior eye segment. It is displayed with a solid line when the instrument is far from the anterior
eye segment. In both cases, the reference is the alignment position in the longitudinal directions.
As the instrument comes near the alignment position, the longitudinal alignment bar grows shorter
gradually.
7
Move the instrument longitudinally with the longitudinal alignment bar as a reference so that the
reflected endothelium image is positioned in the center of the "
" mark on the screen, and
then press the
switch on the joystick.
The image display appears.
L
L
M
S
N
T
C
I
0001
Longitudinal
alignment bar
Photography
3000
2500
2000
1500
1000
T:
ANALYZE DATA EXIT
0001 05/08/05
0.540
13:38
R[C]
Endothelium observation display just before photography
Image
Reflected
endothelium
image