background image

 

3.0 

ASSEMBLY 

 

3.1 

Tools

   

Phillips Screw Driver  
5/32” and 3/16” Allen Wrench 

 

3.2 

Ophthalmic Chair / Stand Position

 

1) 

Place the chair next to the base of the ophthalmic stand or cabinet. Allow it to 
come to room temperature. 

2) 

Remove  the  red  shipping  bolt,  which  secures  the  chair  seat  to  the  base 
assembly (Remove the seat upholstery if installed). 

 

WARNING: Red shipping bolt must be removed before power up.

!

 

3) 

Install  the  upholstery  set  to  the  chair  (if  not  already  installed)  with  the 
instructions included in the upholstery box. 

  

3.3 

Electrical Connections

 

1) 

Connect  the  chair  power  cable  into  the  appropriate  wall  outlet  or  into  the 
ophthalmic stand.  

2) 

Connect the chair control cable to the ophthalmic stand and check the vertical 
movement of the ophthalmic chair. 

 

3.4 

Alternate Connections

 

When  the  Ophthalmic  Chair  model  OC-2200  is  not  used  in  conjunction  with  a 
Topcon  ophthalmic  stand,  use  the  table  below  to  configure  the  electrical  inputs 
correctly. 

 

Stand Cable Wire Code

 

 

Pin 

Number 

Wire Function 

Wire Color 

Ground 

Green 

Neutral 

White 

Hot 

Black 

Up 

Red 

Down 

Violet 

Auto-Return 

Orange 

Relay Engage 

Brown 

Safety Switch 

Yellow 

Safety Switch 

Blue 

 

 

WARNING: 

If the chair is used with a Topcon IS-800 stand, the plastic base shroud (non-fuse 
side) must be removed to make a configuration change on the circuit board. This 
will make the IS-800 safety switch feature operable.

 

 

Summary of Contents for OC-2200

Page 1: ...i...

Page 2: ...ient in a safe condition if a malfunction occurs After Operation Never disconnect any plugs by pulling on the cable Doing so may cause damage to the internal connections If the chair has been idle for...

Page 3: ...ctions 3 3 5 Power Up 4 4 0 OPERATION 4 4 1 Elevation Control 4 4 2 Recline Control 4 4 3 Swivel Brake Operation 4 4 4 Armrest Retraction 4 4 5 Headrest Position 5 4 6 Footrest Retraction 5 5 0 ADJUST...

Page 4: ...rest Height 3 5 89mm 1 5 38mm Full Recline Length 75 1905mm Mode of Operation Continuous Use Type of Protection Class 1 protection Against Electrical Shock Ingress of Water Protection Ordinary Protect...

Page 5: ...es F 23 to 54 Degrees C Relative Humidity Range 30 to 95 Atmospheric Pressure Range 10 1 to 15 4 psi 700 to 1060 hPa Acceptable Storage Time in operating environment 5 years in extreme environment 20...

Page 6: ...appropriate wall outlet or into the ophthalmic stand 2 Connect the chair control cable to the ophthalmic stand and check the vertical movement of the ophthalmic chair 3 4 Alternate Connections When th...

Page 7: ...t of the manual recline ophthalmic chair model OC 2200 is made by pressing the appropriate arrow button on the membrane switch located at the side of the back or with an optional footswitch The chair...

Page 8: ...ng tension Clockwise for more tension counter clockwise for less tension Adjust the clamp tension so that the clamp locking handle closes easily and the pivots at each end of the clamp are secure Repe...

Page 9: ...snap ring off which holds the release button into the membrane switch Remove the membrane switch from the back casting Next remove the switch blank from the opposite side of the back casting and inst...

Page 10: ...ounter clockwise and simultaneously pressing down easily replace the fuses Carefully remove the fuse and check for breaks in the filament or for any corrosion at the ends If in doubt test the fuses wi...

Reviews: