Thank  you  for  purchasing  the  Signature  Series  Model  OC-2200  Ophthalmic  Chair.  Please  read  this 
User's Manual carefully and place it in a convenient location for future reference. 

 

WARNING: The installation of the chair shall be performed only by a qualified

technician.

!

WARNING: The chair is designed for use with TOPCON equipment. If any other

equipment is used, it shall meet the same certifications.

!

WARNING: Any service and/or repairs (other than replacement of the two exter-

nal fuses) shall be performed only by a qualified technician.

!

 

Precautions for Operation 

 

 

Installation: Never position the chair where it will be exposed to moisture, direct sunlight, dust, salty 
air, chemical storage areas, excessive temperature or humidity. Install the chair in a stable place, free 
of uneven floors, vibration and shock. 

 

 

Prior to Operation: Check all cords and/or cabling for proper connections. 

 

 

Operation:  Take  appropriate  action  to  stop  the  chair  from  moving  (leaving  the  patient  in  a  safe 
condition) if a malfunction occurs. 

 

 

After Operation: Never disconnect any plugs by pulling on the cable. Doing so may cause damage to 
the internal connections. If the chair has been idle for a period of time, check all functions and safety 
features for normal operation prior to reuse.  
 

Safety Precautions  

 

 

The OC-2200 exam chair conforms to UL 2601-1 and CAN/CSA-C22.2 No. 601.1-M90. There is no 
disposal of any waste products associated with the OC-2200 exam chair provided that all damaged 
components are returned to Topcon. 

WARNING: Do not load the chair with more than the maximum specified

capacity listed in the specifications section.

!

WARNING: Do not sit on the chair armrests. Doing so may cause damage to

internal chair components.

!

WARNING: If the chair makes a strange noise, discontinue use of the chair

immediately.

!

 

Precautions for Repair 

 

 

When  disassembling,  repairing,  and  reassembling  the  OC-2200,  be  very  careful  not  to  scratch  or 
damage the various mechanical and electrical components of the chair. 
 

 

Always disconnect the power cord when trouble-shooting or repairing electrical system components, 
unless power is absolutely required to perform the repair. 

 

Summary of Contents for OC-2200

Page 1: ...i...

Page 2: ...ient in a safe condition if a malfunction occurs After Operation Never disconnect any plugs by pulling on the cable Doing so may cause damage to the internal connections If the chair has been idle for...

Page 3: ...ctions 3 3 5 Power Up 4 4 0 OPERATION 4 4 1 Elevation Control 4 4 2 Recline Control 4 4 3 Swivel Brake Operation 4 4 4 Armrest Retraction 4 4 5 Headrest Position 5 4 6 Footrest Retraction 5 5 0 ADJUST...

Page 4: ...rest Height 3 5 89mm 1 5 38mm Full Recline Length 75 1905mm Mode of Operation Continuous Use Type of Protection Class 1 protection Against Electrical Shock Ingress of Water Protection Ordinary Protect...

Page 5: ...es F 23 to 54 Degrees C Relative Humidity Range 30 to 95 Atmospheric Pressure Range 10 1 to 15 4 psi 700 to 1060 hPa Acceptable Storage Time in operating environment 5 years in extreme environment 20...

Page 6: ...appropriate wall outlet or into the ophthalmic stand 2 Connect the chair control cable to the ophthalmic stand and check the vertical movement of the ophthalmic chair 3 4 Alternate Connections When th...

Page 7: ...t of the manual recline ophthalmic chair model OC 2200 is made by pressing the appropriate arrow button on the membrane switch located at the side of the back or with an optional footswitch The chair...

Page 8: ...ng tension Clockwise for more tension counter clockwise for less tension Adjust the clamp tension so that the clamp locking handle closes easily and the pivots at each end of the clamp are secure Repe...

Page 9: ...snap ring off which holds the release button into the membrane switch Remove the membrane switch from the back casting Next remove the switch blank from the opposite side of the back casting and inst...

Page 10: ...ounter clockwise and simultaneously pressing down easily replace the fuses Carefully remove the fuse and check for breaks in the filament or for any corrosion at the ends If in doubt test the fuses wi...

Reviews: