25
BASIC OPERATIONS
PATIENT POSITIONING
1
Check the measurement screen.
2
Make sure that the eye height mark is at the center position as explained below.
If the eye height mark is above the center position, press the lower side of the control
panel display, or if it is below the center position, press the upper side of the control panel
display, so as to move the eye height mark to the center position.
3
Have the patient sit in front of the instrument.
4
Adjust the adjustable instrument table or the chair height for the patient to put his/her chin
on the chinrest comfortably.
CAUTION
To avoid electric shock, do not touch the external connection
terminal and the patient at the same time.
CAUTION
To avoid injury when moving the chinrest up/down button, be
careful not to catch the patient's fingers.
CAUTION
To avoid injury when operating the machine, be careful about
the cover not to catch fingers of the patient.
Tell this to the patient, too.
NOTE:
Adjust the height of the adjustable instrument table so that the patient
can sit on the chair comfortably. Otherwise, correct measurement val-
ues may not be obtained.
NOTE:
When operating the instrument, be careful that the instrument does
not touch the patient's lip or nose. If touched, clean the instrument fol-
lowing "CLEANING THE INSTRUMENT" on page 65.
R
R
S
C
A
R1
R1
R2
R2
A
PR1
PR1
R
L
R/K
00001000010000100001
00001000010000100001
0000100001
0000100001
13.75
13.75
dio
dio
AUTO
AUTO
01
01
R
S
C
A
K
R1
R1
R2
R2
A1
A1
PR1
PR1
PR2
PR2
PA1
PA1
PR2
PR2
PA1
PA1
R
S
C
A
K
R1
R1
R2
R2
A1
A1
PR1
PR1
VD
VD
S
ID
A
mm
mm
LOW
LOW
FOG
FOG
0/3
0/3
0/3
0/3
0/3
0/3
0/3
0/3
Center position
Eye height mark
on the center position
Eye height mark