![Topcon BM-7A Instruction Manual Download Page 70](http://html1.mh-extra.com/html/topcon/bm-7a/bm-7a_instruction-manual_1128612070.webp)
4. Communication With a PC
68
4.2.18 RG K Command
The area correction factor data can be read out by this command.
RG[m]K[n]
: m is the area correction factor group number 1 to 10
n is the area number 1 to 5
PC
Command
BM-7A
RG1K3 + (Cr・Lf)
OK + (Cr・Lf)
Area correction (Cr・Lf)
END + (Cr・Lf)
4.2.19 RG L Command
The chromaticity area data to be applied area correction factor are read out by this command.
RG[m]L[n]
: m is the number of the area correction factor group. 1 to 10
n is the area number. 1 to 5
PC
Command
BM-7A
RG1L3 + (Cr・Lf)
OK + (Cr・Lf)
chromaticity area + (Cr・Lf)
END + (Cr・Lf)
4.2.20 WG K Command
The area correction factor can be stored in the instrument.
WG[m]K[n]
: m is the number of the area correction factor group. 1 to 10
n is the area number. 1 to 5
PC
Command
BM-7A
WG1K3‗####‗####‗#### + (Cr・Lf)
OK + (Cr・Lf)
####‗####‗#### + (Cr・Lf)
END + (Cr・Lf)
The values for the correction factors KX, KY and KZ are input in order in ####_####_####.