TH2689/89A Operation manual
53
6.4.3 DISPlay order system
1.
:
DISPlay
:
STATe
?
Function
:
query the currently-displayed function page
Parameter
:
none
Return value
:
{LCTEST | WVTEST | NULL | MAIN | SYSTEM}
2.
:
DISPlay
:
LCTest
Function
:
switch function page as LC TEST
Parameter
:
none
Return value
:
none
3.
:
DISPlay
:
WVTest
Function
:
switch function page as WV TEST
Parameter
:
none
Return value
:
none
6.4.4 LCTest order system
1.
:
LCTest
:
SOURce
:
VOLTage {<numeric_value> | MIN | MAX}
Function
:
set or query the test voltage of LC/IR function
Parameter
:
{<numeric_value> | MIN | MAX}
Return value
:
numeric_value
Instruction
:
MIN
1V
MAX 800V
(
TH2689
)
500V
(
TH2689A
)
2.
:
LCTest
:
SOURce
:
CURRent {<numeric_value> | MIN | MAX}
Function
:
set or query the charge current of LC/IR function
Parameter
:
{<numeric_value> | MIN | MAX}
Return value
:
numeric_value
Instruction
:
MIN
0.5mA
MAX 500.0mA
(
LEV
≤
100V
)
P
MAX
/ LEV
(
P
MAX
=50W
)
3.
:
LCTest
:
CONFigure
:
FUNCtion {SEQ | STEP | CONT}
Function
:
set or query the test mode of LC/IR function
Parameter
:
{SEQ | STEP | CONT}
Return value
:
{SEQ | STEP | CONT}
4.
:
LCTest
:
CONFigure
:
SPEed {FAST | MEDium | SLOW}
Function
:
set or query the test speed of LC/IR function
Parameter
:
{FAST | MEDium | SLOW}
Return value
:
{FAST | MEDium | SLOW}
5.
:
LCTest
:
CONFigure
:
RANGe {<numeric_value> | MIN | MAX}
Function
:
set or query range bin
Parameter
:
{<numeric_value> | MIN | MAX}
Return value
:
numeric_value