DME11D327-06
- 7 -
■
[
P 3 6
] External sensor for the articles to be sterilized calibration
a) Press the temperature setting
▲ ▼ keys to select the system parameter
[
P 3 6
]
b) Press the START key
c) Press the time setting
▲ ▼ keys to set the calibration value for the temperature
sensor inside the chamber.
(e.g., If the temperature measured inside the chamber is 0.8℃ lower than
the temperature displayed on the control panel ( or the set temperature ),
Enter [ -8 ] by pressing the time setting
▲ ▼ keys.)
(e.g., If the temperature measured inside the chamber is 1.1℃ higher than
the temperature displayed on the control panel ( or the set temperature ),
Enter [ +11 ] by pressing the time setting
▲ ▼ keys.)
d) Press the START key to determine the calibration value and store it.
P. 3 6
0
- -l
P. 3 6
0
- -l
The display differs with the type of the product.
P. 3 6
- 8
- -l
Calibration values for the temperature sensor inside the
chamber: -30~+30 (Calibration value 1 = 0.1℃)
P. 3 6
1 1
- -l
Calibration values for the temperature sensor inside the
chamber: -30~+30 (Calibration value 1 = 0.1℃)
P. 3 6
- 8
- -l
P. 3 6
1 1
- -l