4) Connect dummy resistor (123.24 ohm) to CN12 on Main Control PCB
5) Press the START key while more than 2 seconds. It is displayed as shown below [r-2].
6) Connect dummy resistor (151.71 ohm) to CN12 on Main Control PCB
7) Press the START key while more than 2 seconds.
It is displayed as shown first of the display.
8) Press the STOP key or the CHECK key, return to the normal mode.
Temperature display adjustment is updated
.
10. number of lid open/close operation [
c.22
]
1)
Selecting Maintenance code [
C 22
]
2) Press the START key
(e.g., 1,234 times open to close) Lid open to close : 1 time
a.)
TIME
▲ ▼ keys : Times of open/close can be changed
b.)
Press the START key while more than 2 seconds. When the buzzer sounded,
Number of times updated.
c.)
STOP
key : back to maintenance mode
C. 2 2
- - -
- -l
0 0 1
2 3 4
DC
r- 2
0
135.