DEM15D335
- 7 -
8. Setting the manufactureʼs serial number (the last 6
th
– 9
th
digits)
a.) Turn the power switch on while pressing down the FUNCTION , STERILIZE/LIQUID ,and
STERILIZE/NORMAL keys simultaneously to activate the maintenance mode.
b.) Using the temperature setting
▲ and ▼ keys, select the maintenance code [
C17
]
(change in the 6
th
– 9
th
digits of the serial number) and press the START key.
c.) Using the temperature setting ▲ and ▼ keys, set a three-digit sequential serial number (the
6
th
– 8
th
digits of the serial number ) and press the START key.
d.) Using the time setting
▲ and ▼ keys, set a one digit serial number (the 9
th
digit of the
serial number ) and press the START key.
When a manufacturerʼs serial number consists of eight digits, having a sequential serial number
of only three digits, the last 9
th
digit is set to “
–
“.
e.) Press the STOP key.
9. Setting the Print date and time
a.) Press and hold down the FUNCTION key for approximately two seconds while the ready lamp is
illuminated.
“
F0 1
”
will start flashing in the temperature display.
b.) Press the temperature
setting
▲ and ▼ keys until the code of a desired setting item shows in
the display. The time display will show the code of a desired setting item.
CODE
SETTING ITEM
SETTING RANGE
F09
YEAR
00-99
F 10
MONTH
01-12
F 1 1
DAY
01-31
F 12
HOUR
00-23
F 13
MINUTE
00-59
* To set the year, enter the last two digits of the year.
c.) Press the START key. The time display will start flashing.
d.) Press the time setting ▲ and ▼ keys until the display
shows a desired set value.
e.) Press the START key again to enter the value.
*When the set value for the minutes is entered, seconds are set as 00.