DEM15D335
- 5 -
5. Pressure display and adjustment
1) Tur
n on the power switch while pressing the FUNCTION , STERILIZE/LIQUID and
STERILIZE/NORMAL keys at the same time.
As you make an adjustment of pressure, the lid opens before the maintenance mode.
2) Selecting Maintenance code [
C 02
]
3) Confirm that the lid opens before and no pressure in the chamber
And Press the START key while more than 2 seconds. When the buzzer sounded,
Offset is updated
.
4) Press the STOP key or the CHECK key, return to the normal mode.