Tomy SX-300 Installation Manual Download Page 4

DEM15D335

   

 

 

- 3 -

 

3. Remove the control PCB and install control PCB 

 

 

1)Remove the rear panel. 
2)Disconnect each wiring harness and remove the control PCB. 
3)

Put in the Dot-impact printer programed Control PCB with Option PCB and   

reconnect wires same as before. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 

Option PCB 

Control PCB TH200901

Thermal 

printer 

programed 

Control 

PCB 

Exchange 

Summary of Contents for SX-300

Page 1: ...ol PCB TH 200901 Autoclave SX 300 500 700 SX 300E 500E 700E TOMY SEIKO CO LTD TOMY DIGITAL BIOLOGY CO LTD 3 14 17 Tagara Nerima ku Tokyo 179 0073 Japan e mail info digital biology co jp URL http www digital biology co jp phone 81 3 5971 8160 fax 81 3 3970 6036 ...

Page 2: ... Install Printer ASSY onto Front Panel ASSY Mount screws 4 places from back of Front panel 3 Connect the wiring to the printer 4 attach attached aluminum glass cloth tape to the following position Attach to the square hole on the printer ASSY in order not to water entering Front Panel ASSY Front of Front Panel ASSY Printer ASSY Screws 4 places Back of Front panel ASSY Tape attaching ...

Page 3: ...l the panel out to remove it 3 Remove the exhaust bottle cover located behind the exhaust bottle Unscrew the four screws securing the cover 4 Install the back plate on the exhaust bottle cover removed Install and secure the back plate using cross recessed pan head screw M4 x 8 5 Install the exhaust bottle cover into the autoclave and put the exhaust bottle back in the place Front panel Exhaust bot...

Page 4: ...Remove the rear panel 2 Disconnect each wiring harness and remove the control PCB 3 Put in the Dot impact printer programed Control PCB with Option PCB and reconnect wires same as before Option PCB Control PCB TH200901 Thermal printer programed Control PCB Exchange ...

Page 5: ...om plate then pass the harness through the lifter Secure the wiring harness along with a power supply line harness using a push mount tie to the electric chassis Run the wire connected to the option PCB along the exhaust bottle cover and fan cover and then secure it with the junction harness using the push mount tie Wire connected to the terminal block Wire connected to the option PCB Ground fixin...

Page 6: ...in the chamber And Press the START key while more than 2 seconds When the buzzer sounded Offset is updated 4 Press the STOP key or the CHECK key return to the normal mode Offset is updated Destination of the ground wire from the printerʼs front panel Wire connected to the terminal block Wire connected to the option PCB Temperature display Time display Warming Timer display FUNCTION key STERILIZE L...

Page 7: ...nd hold down the START key e Press the STOP key 7 Setting the manufactureʼs serial number the first five or les digits a Turn the power switch on while pressing down the FUNCTION STERILIZE LIQUID and STERILIZE NORMAL keys simultaneously to activate the maintenance mode b Using the temperature setting and keys select the maintenance code C16 change in the first five digits of the serial number and ...

Page 8: ...f eight digits having a sequential serial number of only three digits the last 9th digit is set to e Press the STOP key 9 Setting the Print date and time a Press and hold down the FUNCTION key for approximately two seconds while the ready lamp is illuminated F0 1 will start flashing in the temperature display b Press the temperature setting and keys until the code of a desired setting item shows i...

Page 9: ...tion and check that the printer prints out the operating result Reference The illustration on the right indicates a printed image b Make sure that the serial number date and time set according to the procedure 4 are correct 11 Completing and returning the validation report to TOMY ...

Page 10: ... All work must be done according to the work procedure instruction sheet Make sure that the product functions normally Validation Report To TOMY Digital Biology Co Ltd I hereby represent that we have validated the custom product for its compliance with the custom details described above at user s site and report the validation result Model name Serial number year month day Approved by Validated by...

Reviews: