3.2.3 CONTROLLER OPERATION
Following are the controller panel indications possible under various system conditions:
SLP = Standby Mode
XXX = Product water conductivity (press C/T button to read cond.)
XX.X = Product water temperature in degrees C (press C/T button to read temp.)
FULL = Product water tank full
ALA + HIGH PRESS LED RED = High pressure pump output over limit (optional feature)
ALA + LOW PRESS LED RED
=
Feed water pressure under limit
ALA + LOW FEED PRESS LED RED = Feed water storage tank level low (optional feature)
ALA + HIGH LED RED = Product water conductivity over limit
Green LED displays on the controller panel indicate that the parameter is within normal limits or that
the element (pump or valve) is active. Flush mode settings may need to be set initially and after any
power interuption to the controller. To change settings on the controller proceed as follows:
1. Press the SET button. The conductivity calibration factor will appear. This value has been
factory set and should not be changed.
2. Press the SET button a second time. The value displayed is the high limit for product water
conductivity (in microseimens). Readings above this value will result in an audIble alarm as well as
an ALA display and red HIGH LED indication. To change the limit value use the arrow buttons to go
up or down. When the value is as desired press the C/T button to save the new value.
2. Press SET a third time. The display will now indicate the duration of flush that will occur when-
ever the system starts as “aXXX”. The range is from 000 to 249 seconds (4 minutes). If no startup
flush is desired or if the flush option is not installed set the value to 001 otherwise set the time to the
desired duration. When done, press the C/T button to save the value.
3. Press SET a fourth time. The display will now indicate the duration of flush that will occur
whenever the system shuts down as “bXXX”. The range is from 000 to 249 seconds (4 minutes). If
no shut down flush is desired or if the flush option is not installed set the value to 001 otherwise set
the time to the desired duration. When done, press the C/T button to save the value.
4. Press the SET button a fifth time. The number now displayed as “C-XX” indicates the interval in
continuous running that the system will stop and automatically go into a shut down flush, and then
resume running after the flush is complete. The value range is 00 to 99 hours. If no automatic flush
is desired, set the value to 99 hours otherwise set to the desired interval. Again, this feature is only
active if the flush option is installed. If not, set the value to 99. Press the C/T button to save the
value. Pressing the SET button one last time will return the display to normal indication.
3.2.4 OPERATING PARAMETER CONDITIONS
MODEL NO.
TV-6000
TV-8000
Membrane Type
Low Energy TF
Low Energy TF
Feed Pressure
25 psi (1.7 bar)
25 psi (1.7 bar)
Min. Feed Flow Rate (without recirculation) 8.2GPM @ 60Hz 13.6 GPM @ 60Hz
Maximum Temperature (Feed)
113°F (45°C)
113°F (45°C)
Maximum Chlorine (continuous)
<0.1 ppm
<0.1 ppm
Maximum Total Dissolved Solids
5,000 ppm
5,000 ppm
pH Range
3.0 - 10.0
3.0 - 10.0
Iron
0.0 ppm
0.0 ppm
Hydrogen Sulfide
0.0 ppm
0.0 ppm
10
Summary of Contents for TV-6000
Page 2: ...TOMAR TV 6000 SYSTEM ...
Page 7: ...Figure 1 Typical System Configuration TV 12K shown 5 ...
Page 13: ...Figure 2 System Plumbing Diagram 11 ...
Page 14: ...Figure 3 System Electrical Schematic Diagram ...
Page 20: ...Figure 4 PVC Vessel Membrane Assembly 18 ...
Page 21: ...Figure 5 Stainless Steel Vessel Membrane Assembly 19 ...
Page 28: ...25 NOTES ...
Page 31: ......