28
DIGITAL TORQUE WRENCH “DATA TORK” MODEL
CEM3-G
Connect the BC-3-G charger to the CEM3 charge jack. Make sure BC-3-G charger
is connected to the power source. Green lamp on BC-3-G turns on when charging is complete
(it takes about 3.5 hours from the empty condition).
(1) Check the voltage on the charger and use the appropriate power source.
(2) Stop charging as soon as the green lamp on the charger turns on. Excessive charging may
shorten the battery life.
(3) The product cannot operate when it is connected to the charger.
(4) If the green lamp on the charger turns on and the red lamp starts to blink, it indicates an
error. Stop using it immediately, and contact Tohnichi or nearest Tohnichi distributor.
(5) Temperature must be kept within 0-40 degree Celsius range when charging.
(6) If it should emit some abnormal smell or generates abnormal heat, stop using it immediately
and move it to a safe place. Contact Tohnichi or nearest Tohnichi distributor.
(7) When not in use for a long time, charge it to full, and remove the battery to keep it. It is
recommended that it should be charged at least once every half a year.
After charging, there is no need to push the reset button.
13
Charging
Warnings
Error Message
Indication
Err1
key is continuously pushed.
Err2
key is continuously pushed.
Err3
key is continuously pushed.
Err4
key is continuously pushed.
Err5
key is continuously pushed.
Error Message
Turn off the
once and turn it on without touching
any other keys.
If Err disappears, it should work properly.
If Error does not disappear, it needs to be repaired.
Please contact TOHNICHI or your nearest distributor.
Solution
It needs to be repaired. Please contact TOHNICHI
or your nearest distributor.
Malfunction of the torque sensor or circuit board.
Push
key at no loading condition.
If Err 9 disappears, it should work properly.
If Err 9 remains, it needs to be repaired. Please
contact TOHNICHI or your nearest distributors.
Err9