![TMI Products TQC Sheen 32-76e Series Manual Download Page 35](http://html1.mh-extra.com/html/tmi-products/tqc-sheen-32-76e-series/tqc-sheen-32-76e-series_manual_1121874035.webp)
32-76-00-0002
35
Rev. 6/06/18
•
Enter the load cell’s capacity in Newtons.
•
Press the checkmark to confirm.
•
Enter the Load Cell’s sensitivity in millivolts per Volt.
•
This number is a characteristic of the load cell from its original manufacturer. TMI ships the OEM
certificate and prints the sensitivity on the connector inside the load cell arm case. If none of these
are available you can enter 3.0 here and then you must do a full calibration.
•
Press the checkmark to confirm.
•
A partial calibration takes values from a previous Full calibration and a load cell’s known
sensitivity to calculate gain values.
•
Press the “Full” button if you have calibrated weights and the hanging fixture.
Coefficient of Friction - Peel Tester-10262018-ab5700-m44.pdf 35
Coefficient of Friction - Peel Tester-10262018-ab5700-m44.pdf 35
15-6-2020 13:41:02
15-6-2020 13:41:02