TLV VFM-T2 Instruction Manual Download Page 18

 

172-65256M-00  (VFM-T2)  7 Sep 2001

 

17

Operation 

Preparations Before Operation 

Inspect the counter and associated instruments to ensure that they are correctly 
installed and wired and that nothing has been overlooked. 

 

<Warning> Use of a power source of an incorrect voltage may result in 
burnout of the counter. Make certain to use a power source of the rated 
voltage. 

 
(Pre-operational Check) 
Supply power to the counter. Press the [MODE] button to test function and check for 
irregularities. 

 

<Caution> Avoid repeatedly turning the power ON and OFF, as this may 
cause the counter to begin counting. 

Next, supply a simulated pulse train of the appropriate type to confirm that the counter 
counts and displays a correct totalized flow and instantaneous flow rate in response. 
 

Operation 

After pre-operational preparations are completed, turn on the power. 
 
 

Troubleshooting 

u

 IMPORTANT 

u

 

When internal trouble is suspected, contact TLV after inspecting the 
counter by following the steps below. 

Symptom 

Check Item 

Possible Causes 

Is power being supplied? 
(1) Check the voltage 
(2) Inspect the fuse 

(1) The supply voltage is incorrect 
(2) The fuse is blown (internal) 
(3) The power supply unit is   
faulty 

The display 
indicates neither 
total flow nor 
instantaneous flow 
rate 

Is a pulse input train 
being supplied? 
(1) Is the wiring correct? 
(2) Is the pulse input train   
reaching the counter? 

(1) The input lines are wired incorrectly 
(2) The transmitter is faulty 
(3) The waveform shaping circuit is   
faulty 
(4) The custom (

µ

 CPU) is defective 

u

 IMPORTANT 

u

 

If the problem is suspected to be other than one listed above, contact the 
TLV service network. When contacting TLV, be prepared to provide details 
such as the product name, model and symptoms of the problem. 

Summary of Contents for VFM-T2

Page 1: ...65256M 00 VFM T2 7 September 2001 Vortex Flowmeter Counter VFM T2 Copyright 2001 by TLV CO LTD All rights reserved ISO 9001 ISO 14001 Manufacturer Kakogawa Japan is approved by LRQA LTD to ISO 9001 14...

Page 2: ...g Terminal Block 7 Input Signal Connection 7 Configuration Block Diagram 8 Functions and Operation 9 The LCD Counter Display 9 Verifying the Meter Factors 9 Calculating Total and Instantaneous Flow Ra...

Page 3: ...pearance to make sure nothing is out of the ordinary Also be sure to read this manual carefully before use and follow the instructions to be sure of using the product properly If detailed instructions...

Page 4: ...r damage occurring as a result of failure to observe these precautions Symbols Indicates a DANGER WARNING or CAUTION item DANGER Indicates an urgent situation which poses a threat of death or serious...

Page 5: ...rt where at all possible transport the product to the installation site in the original packaging in which it was shipped from TLV 2 During transport avoid subjecting the product to strong impacts or...

Page 6: ...croprocessor and is lightweight and compact in design It offers four independent functions cumulative total flow instantaneous flow rate hourly and per minute total flow and resettable total in a sing...

Page 7: ...o injury fires damage or other accidents may occur Do not touch the terminals during transmission Failure to observe this precaution could result in such hazards as damage to the product malfunctions...

Page 8: ...m Caution As incorrect combinations could lead to serious damage to these units be sure to check again after connecting the cables Description of Field Wiring Terminals Name Terminal No Connection and...

Page 9: ...cuit in the next step When the full scale frequency is in the range of 2 4Hz and the waveform ratio of the input pulse is 1 1 J1 CLOSE F I Circuit converts an input frequency of 4 20mA or 1 5V to an a...

Page 10: ...sequence Verifying the Meter Factors If the MODE button is pushed when the select switch is set to either C or 4 the meter factors are displayed in the following sequence NOTE For items marked with a...

Page 11: ...rate in full scale is held within 1 100 T 200ms NOTE It is designed so that when T 5000 ms the instantaneous flow rate on the display will show 0 NOTE Set it so that sampling cycles input period 5000...

Page 12: ...ove the screws from the rear plate and then take the rear plate off 3 Pull the internal unit out to the rear Waveform Shaping Function Specifications for Type of Pulse Input and Input Settings Jumper...

Page 13: ...ith the individual specifications Setting the Pulse Output Specifications Selection Jumper J3 1 Synchronized input 1 1 A C 2 Scaled display units standard B C Power Board underside Setting Hard Dividi...

Page 14: ...manner only the procedure for the meter factor is explained here to avoid duplication Ex Changing the meter factor from 1 2345 to 0 9876 1 Change the position of the select switch from the default pos...

Page 15: ...3 Verifying the New Meter Factor Setting Change the select switch from F to C and press the MODE button to confirm each individual digit The new meter factor Ex is displayed 14 Change the select switc...

Page 16: ...fixed l l Set the pulse width to a suitable flow rate range to ensure that the pulse does not become too heavy Operation and Adjustment for Setting and Modifying the Analog Output Circuit Optional The...

Page 17: ...09 9 B 110 0 119 9 C 120 0 129 9 D 130 0 139 9 E 140 0 149 9 F ON ON ON ON 150 0 199 9 2 OFF OFF OFF OFF 200 0 299 9 3 OFF 300 0 399 9 4 ON 400 0 499 9 5 500 599 6 600 699 7 700 799 8 800 899 9 900 99...

Page 18: ...onse Operation After pre operational preparations are completed turn on the power Troubleshooting u IMPORTANT u When internal trouble is suspected contact TLV after inspecting the counter by following...

Page 19: ...to 1 10 or 1 100 When scaler value exceeds 1 150Hz maximum Selection Scaled same unit as displayed standard or synchronized input unscaled Output Signal Open collector after optical isolation Capacity...

Page 20: ...improper installation use handling etc by other than TLV CO LTD authorized service representatives Malfunctions due to dirt scale rust etc Malfunctions due to improper disassembly and reassembly or i...

Reviews: