TLV VFM-T2 Instruction Manual Download Page 11

 

172-65256M-00  (VFM-T2)  7 Sep 2001

 

10

 

Calculating Total and Instantaneous Flow Rates 

(1) Total: both cumulative and resettable total 
 
Input Pulses 

  Frequency  

  Meter Factor  

  Frequency   

 To Individual   

 
 
 
 
 
(2) Instantaneous: used only when input pulses have very small periodic variation. 

The period of sample cycles is measured down to the unit of 1 ms, calculated 
using the following equation and the result shown on the display. 

 

3600 

×

 A 

×

 F 

Flow rate per hour = b1 = 

T (ms) 

×

10 

NOTE: A: number of sampling cycles 

F: meter factor

 

 

   

T: interval of input pulse 

b1 

 

Flow rate per minute = b2 = 

60 

×

100 

 

 
NOTE: The number of sampling cycles is determined such that the accuracy of the 

instantaneous flow rate in full scale is held within 

±

 1% (100 

 T 

 

200ms). 

NOTE: It is designed so that when T 

 

5000 (ms), the instantaneous flow rate on the 

display will show “0”. 

NOTE: Set it so that sampling cycles × input period 

 5000 (ms). 

NOTE: Note that when “hard” dividing, the units for instantaneous flow rate are 

modified. 

 

Setting and Adjusting Jumpers, Switches and Volumes 

l

 

The multifunction counter is preset to the customer specifications at the 

factory before shipment. Unless it is necessary to change the preset 
settings in the field, further adjustment should not be required. 

 
NOTE: It is necessary to remove the internal components (motherboard) of the unit 

when setting, modifying or adjusting jumpers, switches or volumes. 

 
 

Hard 

Soft 

1/1,1/10 or 1/100 

0.0001 - 1.9999 

(1.000 set for input 

of unit pulse) 

1/1,1/10 or 1/100 

µ

 - CPU 

Reduction 
(Dividing)   

Counter and 
Output Unit 

Reduction 
(Dividing)   

Summary of Contents for VFM-T2

Page 1: ...65256M 00 VFM T2 7 September 2001 Vortex Flowmeter Counter VFM T2 Copyright 2001 by TLV CO LTD All rights reserved ISO 9001 ISO 14001 Manufacturer Kakogawa Japan is approved by LRQA LTD to ISO 9001 14...

Page 2: ...g Terminal Block 7 Input Signal Connection 7 Configuration Block Diagram 8 Functions and Operation 9 The LCD Counter Display 9 Verifying the Meter Factors 9 Calculating Total and Instantaneous Flow Ra...

Page 3: ...pearance to make sure nothing is out of the ordinary Also be sure to read this manual carefully before use and follow the instructions to be sure of using the product properly If detailed instructions...

Page 4: ...r damage occurring as a result of failure to observe these precautions Symbols Indicates a DANGER WARNING or CAUTION item DANGER Indicates an urgent situation which poses a threat of death or serious...

Page 5: ...rt where at all possible transport the product to the installation site in the original packaging in which it was shipped from TLV 2 During transport avoid subjecting the product to strong impacts or...

Page 6: ...croprocessor and is lightweight and compact in design It offers four independent functions cumulative total flow instantaneous flow rate hourly and per minute total flow and resettable total in a sing...

Page 7: ...o injury fires damage or other accidents may occur Do not touch the terminals during transmission Failure to observe this precaution could result in such hazards as damage to the product malfunctions...

Page 8: ...m Caution As incorrect combinations could lead to serious damage to these units be sure to check again after connecting the cables Description of Field Wiring Terminals Name Terminal No Connection and...

Page 9: ...cuit in the next step When the full scale frequency is in the range of 2 4Hz and the waveform ratio of the input pulse is 1 1 J1 CLOSE F I Circuit converts an input frequency of 4 20mA or 1 5V to an a...

Page 10: ...sequence Verifying the Meter Factors If the MODE button is pushed when the select switch is set to either C or 4 the meter factors are displayed in the following sequence NOTE For items marked with a...

Page 11: ...rate in full scale is held within 1 100 T 200ms NOTE It is designed so that when T 5000 ms the instantaneous flow rate on the display will show 0 NOTE Set it so that sampling cycles input period 5000...

Page 12: ...ove the screws from the rear plate and then take the rear plate off 3 Pull the internal unit out to the rear Waveform Shaping Function Specifications for Type of Pulse Input and Input Settings Jumper...

Page 13: ...ith the individual specifications Setting the Pulse Output Specifications Selection Jumper J3 1 Synchronized input 1 1 A C 2 Scaled display units standard B C Power Board underside Setting Hard Dividi...

Page 14: ...manner only the procedure for the meter factor is explained here to avoid duplication Ex Changing the meter factor from 1 2345 to 0 9876 1 Change the position of the select switch from the default pos...

Page 15: ...3 Verifying the New Meter Factor Setting Change the select switch from F to C and press the MODE button to confirm each individual digit The new meter factor Ex is displayed 14 Change the select switc...

Page 16: ...fixed l l Set the pulse width to a suitable flow rate range to ensure that the pulse does not become too heavy Operation and Adjustment for Setting and Modifying the Analog Output Circuit Optional The...

Page 17: ...09 9 B 110 0 119 9 C 120 0 129 9 D 130 0 139 9 E 140 0 149 9 F ON ON ON ON 150 0 199 9 2 OFF OFF OFF OFF 200 0 299 9 3 OFF 300 0 399 9 4 ON 400 0 499 9 5 500 599 6 600 699 7 700 799 8 800 899 9 900 99...

Page 18: ...onse Operation After pre operational preparations are completed turn on the power Troubleshooting u IMPORTANT u When internal trouble is suspected contact TLV after inspecting the counter by following...

Page 19: ...to 1 10 or 1 100 When scaler value exceeds 1 150Hz maximum Selection Scaled same unit as displayed standard or synchronized input unscaled Output Signal Open collector after optical isolation Capacity...

Page 20: ...improper installation use handling etc by other than TLV CO LTD authorized service representatives Malfunctions due to dirt scale rust etc Malfunctions due to improper disassembly and reassembly or i...

Reviews: