172-65592MA-07 (TM5N/TM5N-EX TrapMan Type P) 12 May 2022
28
6.4 Messages Displayed During Measurement
L
? ? ? ? ?
?
H
TEMP. 150
C 15s
The intensity of the detection level will
be displayed in bar graph.
15 second
measurement count
Surface temperature
L
? ? ? ? ?
?
H
TEMP. 302
F 15s
The intensity of the detection level will
be displayed in bar graph.
15 second
measurement count
Surface temperature
6.5 Entering the Pressure
1. When measurement ends, the screen that allows the user to
enter the trap inlet pressure (gauge pressure) will
automatically appear. The pressure displayed in this screen
is the saturation steam pressure equivalent to the measured
surface temperature unless measurements have been done
in the past for that control number; then the pressure entered
in the past will be displayed.
2. If you know the operating pressure directly ahead of the inlet
for the measured trap, enter that value on the numeric keys.
For example, if the pressure is 5.5 kg/cm
2
G, enter
0
5
5
. The decimal point position is fixed, so it
need not be entered.
For example, if the pressure is 78 psig, enter
0
7
8
. There is no allowance for fractional
pressures (no decimal).
Press
ENT
to establish the value you have entered.
No. A01-00000
PRESS. 01.5 KG
No. A01-00000
PRESS. 021 psi
No. A01-00000
PRESS. 05.5 KG
No. A01-00000
PRESS. 078 psi