Recommended separation distance
between portable and mobile RF
communications equipment and ClearUP
ClearUP® is intended for use in the
electromagnetic environment in which radiated
RF disturbances are controlled. The customers
or the users of ClearUP can help prevent
electromagnetic interference by maintaining a
minimum distance between portable and mobile
RF communications equipment (transmitters)
and ClearUP as recommended below,
according the maximum output power of the
communications equipment.
For transmitters rated at a maximum output power not
listed above, the recommended separation distance
d
in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Separation distance according the
frequency of transmitter in meter
Output
Power of
Transmitter
in Watt
150 kHz to 80
MHz
d= 1.2 √P
80 MHz to
800MHz
d= 1.2√P
800MHz
to 2.5GHz
d= 1.2√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.78
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23